1630-2/1630-2 FC
Calibration Manual
10
4.
Push
A
to cancel calibration and go back to the measurement mode.
To confirm calibration points:
1.
At each calibration step, apply the input signal according to the indication on the lower portion of the screen.
Table 6 is a list of the calibration points:
• 8 calibration points for ac leakage current verification and adjustment
• 15 calibration points for
Ω
verification and adjustment
If the screen shows "- - - -", do not apply any signal to jaw, attach the Clamp to a conductor, or open the clamp.