1630-2/1630-2 FC
Calibration Manual
14
Finish Calibration:
When calibration is successful, the Clamp exits the calibration mode and goes to the measurement mode.
Table 6. Calibration Points
Step
Display
Calibrator Output Value
1
CAL 01
---
mA AC
2
CAL 02
0.200
0.100
mA
60 HZ
3
CAL 03
0.400
0.200
4
CAL 04
4.000
2.000
5
CAL 05
40.00
20.00
6
CAL 06
400
200.0
7
CAL 07
4000
2000
8
CAL 08
40.00
A AC
20.00
A
9
CAL 09
---
Ω
10
CAL 10
0.100
0.100
Ω
11
CAL 11
0.250
0.250
12
CAL 12
0.500
0.500
13
CAL 13
1.000
1.000
14
CAL 14
2.000
2.000
15
CAL 15
5.000
5.000
16
CAL 16
10.00
10.00
17
CAL 17
20.00
20.00
18
CAL 18
50.00
50.00
19
CAL 19
100
100.0
20
CAL 20
200
200.0
21
CAL 21
400
400
22
CAL 22
600
600
23
CAL 23
1000
1000