![Fluke 1577 Calibration Manual Download Page 27](http://html1.mh-extra.com/html/fluke/1577/1577_calibration-manual_2309485027.webp)
Insulation Multimeters
Performance Test
17
Table 3. DMM Accuracy Tests (cont)
UUT Display Reading
Step
UUT
Function
UUT
Range
UUT
Input Level
Fluke 1577
Fluke 1587
16
V dc
600.0V
-600.0V
-601.4 to -598.6
-600.3 to -599.7
17
V dc
1000V
1000V
996 to 1004
997 to 1003
18
mV dc
600.0 mV
3 mV
2.9 to 3.1
2.9 to 3.1
19
mV dc
600.0 mV
30 mV
29.9 to 30.1
29.9 to 30.1
20
mV dc
600.0 mV
-600 mV
-601.3 to 598.7
-600.7 to -599.3
21 Ohms
600.0
Ω
0
-0.2 to 0.2
-0.2 to 0.2
22 Ohms
600.0
Ω
600
Ω
592.6 to 607.4
594.4 to 605.6
23 Ohms
6.000
k
Ω
6000
Ω
5.926 to 6.074
5.944 to 6.056
24 Ohms
60.00
k
Ω
60.00
k
Ω
59.26 to 60.74
59.44 to 60.56
25 Ohms
600.0
k
Ω
600.0
k
Ω
592.6 to 607.4
594.4 to 605.6
26 Ohms
6.000
M
Ω
6.000
M
Ω
5.926 to 6.074
5.944 to 6.056
27 Ohms
50.00
M
Ω
50.00
M
Ω
48.97 to 51.03
49.22 to 50.78
28
Capacitance
1000 nF
700 nF
n/a
690 to 710
29
Capacitance
9999 uF
8000 uF
n/a
7814 to 8186
30
Continuity
n/a
103
Beeper off
Beeper off
31
Continuity
n/a
23
Beeper on
Beeper on
32
Diode
n/a
2.5 V
n/a
2.447 to 2.553
33
mA ac
60.00
3 mA, 45Hz
2.92 to 3.08
2.93 to 3.07
34
mA ac
600.0
400 mA, 60Hz
391.8 to 408.2
393.8 to 406.2
35
mA dc
60.00
60.00 mA
59.38 to 60.62
59.86 to 60.14
36
mA dc
600.0
400.0 mA DC
395.8 to 404.8
399 to 401
* To ensure accurate frequency use the Calibrator Frequency mode.