Insulation Multimeters
Required Tools and Equipment
15
Reassembling the Meter
To reassemble the Meter, logically reverse the previous disassembly procedures. In
the process, make sure to re-establish all electrical connections; specifically, the
elastomeric strip for the LCD, the red and black plastic shields for the input terminals,
and the two spring contacts for the Insulation Test Probe. Also make sure all parts are
correctly aligned and positioned; do not force-fit any of the parts into position.
Required Tools and Equipment
Table 1 lists the require equipment for performing the Performance Test and
Calibration adjustments. If a recommended model is not available, use a substitute
with equivalent or better specifications.
Table 2. Required Tools and Equipment
Equipment Required
Characteristics
Recommended
Model
Calibrator
AC Voltage Range: 0-1000V
Accuracy: +/- 0.25%
Frequency Range: 50Hz - 5 kHz
Accuracy: +/-3%
DC Voltage: 0-1000V
Accuracy: +/-0.075%
Current Range: 0-600 mA
Accuracy: AC mA = +/-0.38%
Frequency: 45Hz -1 kHz
DC mA =+/-0.058%
Frequency Source: 100Hz - 100 kHz
Accuracy: +/-0.028%
Amplitude: 5 Volt
Accuracy: +/-5%
Ohms Range: 0-50 M
Ω
Accuracy: +/-0.23%
Capacitance Range: 700 nF - 8000 uF
Accuracy: +/-0.35%
Insulation Resistance: 500 k
Ω
- 1.7 G
Ω
Accuracy: +/-0.9%
Fluke 9100 w/135 option
DMM (Test DMM)
DC Volts: 0 to 1000V
Accuracy: +/-5%
DC Current: 0 - 2mA
Accuracy: +/-1%
Fluke 189
Power Supply
DC Volts: 0 to 15V
Generic Lab Supply
HV Probe (Divider)
DC Voltage Range: 1kV -5 kV
Accuracy: 1%
Division ratio: 1000:1
Input resistance: 1000 M
Ω
Fluke 80k-6
Thermocouple
K Type with Dual Banana connector
Fluke 80BK
Thermocouple Mini-
Connector
K Type
Fluke 80CK-M