1587/1577
Calibration
2
Contacting Fluke
To contact Fluke, call one of the following telephone numbers:
•
Technical Support USA: 1-800-44-FLUKE (1-800-443-5853)
•
Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853)
•
Canada: 1-800-36-FLUKE (1-800-363-5853)
•
Europe: +31 402-675-200
•
Japan: +81-3-3434-0181
•
Singapore: +65-738-5655
•
Anywhere in the world: +1-425-446-5500
Or, visit Fluke's website at
www.fluke.com
.
To register your product, visit
http://register.fluke.com
.
To view, print, or download the latest manual supplement, visit
http://us.fluke.com/usen/support/manuals
.
General Specifications
Maximum Voltage Applied to any Terminal ........ 1000 V ac rms or dc
Storage Temperature.......................................... -40
°
C to 60
°
C (-40
°
F to 140
°
F)
Operating Temperature ...................................... -20
°
C to 55
°
C (-4
°
F to 131
°
F)
Temperature Coefficient ..................................... 0.05 x (specified accuracy) per
°
C for temperatures <18
°
C or >28
°
C (< 64
°
F or > 82
°
F)
Relative Humidity................................................ Noncondensing
0 % to 95 % @ 10
°
C to 30
°
C (50
°
F to 86
°
F)
0 % to 75 % @ 30
°
C to 40
°
C (86
°
F to 104
°
F)
0 % to 40 % @ 40
°
C to 55
°
C (104
°
F to 131
°
F)
Vibration.............................................................. Random, 2 g, 5-500 Hz per MIL-PRF-28800F, Class 2 instrument
Shock .................................................................. 1 meter drop per IEC 61010-1 2
nd
Edition (1 meter drop test, six
sides, oak floor)
Electromagnetic Compatibility ............................ In an RF field of 3 V/M, accuracy = specified accuracy except in
temperature: accuracy = specified accuracy
±
5
°
C (9
°
F). (EN
61326-1:1997).
Safety.................................................................. Complies with ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA-
C22.2 NO. 61010-1-04, and IEC/EN 61010-1 2
nd
Edition for
measurement category III 1000 V (CAT III) and
CAT IV 600 V.
Certifications ....................................................... CSA per standard CSA/CAN C22.2 No. 61010.1-04; TUV per
standard EN 61010
Part 1-1002
Batteries.............................................................. Four AA batteries (NEDA 15A or IEC LR6)
Battery Life.......................................................... Meter use 1000 hours; Insulation test use: Meter can perform at
least 1000 insulation tests with fresh alkaline batteries at room
temperature. These are standard tests of 1000 V into 1 M
Ω
with a
duty cycle of 5 seconds on and 25 seconds off.
Size ..................................................................... 5.0 cm H x 10.0 cm W x 20.3 cm L
(1.97 in H x 3.94 in W x 8.00 in L)
Weight................................................................. 550 g (1.2 lb.)
IP Rating ............................................................. IP40
Altitude
Operating ........................................................ 2000 m CAT III 1000 V, CAT IV 600 V; 3000 m CAT II 1000 V,
CAT III 600 V
Storage ........................................................... 12,000 m
Over-Range Capability
.......................................
110% of range except for capacitance which is 1 %
Compliance to EN 61557.................................... IEC61557-1, IEC61557-2