10.4
Power Control Functions
The digital interface is capable of controlling the heating functions so that the
bath can be remotely operated at any temperature within the range of the bath.
To allow the interface to control the heating, the front panel controls are dis-
abled by switching the heater switch to “LOW”. Otherwise, the interface can
not switch the heater functions off. The 6020 series baths have one control
function with the digital interface. This control function is for the heater
settings.
To control the heater with the digital interface, the front panel heater switch
must be set to “LOW”. The heater function is controlled with the “F1" com-
mand. This command is either set to ”0" or “1" according to Table 4. The de-
fault setting is ”F1=0" which is “LOW”. Sending a command without a
parameter returns the state, “1" for on or ”0" for off.
57
10 Digital Communication Interface
Power Control Functions
Power
F1
LOW
0
HIGH
1
Table 4
Power Control Functions