Insert the probe into the 1/4 inch probe hole at the top left side of the bath lid.
The tip of the probe must be well immersed in the fluid. The probe connector is
plugged into the rear of the bath into the socket labeled “PROBE”.
5.5
Power
With the bath power switch off, plug the bath into an AC mains outlet of the
appropriate voltage, frequency, and current capacity. See Section
3.1
,
Specifications.
Be sure the stirring motor power cord is plugged into the “STIRRER” socket at
the back of the bath.
5.6
Tipping Prevention Brackets Installation (6024
with model 2072 cart only)
WARNING:
Hart Scientific recommends installing the bath on a heat-re-
sistant non-flammable surface.
In order to create the safest possible conditions under use, the bath comes
equipped with Tipping Prevention Brackets. The installation of the brackets is
required for compliance with the
International Safety Standard IEC 61010-1,
Section 7.3, Stability,
which applies to the stability of the bath under normal
operating conditions.
5.6.1
Installation on a Concrete Floor
Using a concrete drill and concrete drill bit, drill three 1/4 inch x 1 1/4 inch
deep (approximately 6.5 mm x 32 mm) holes in the concrete floor using the
bracket to mark the hole placement. Drop the flare anchor bolt into the hole.
Tightening the screw expands the anchor in the drilled hole and secures the
bracket. Ensure that the bracket is installed in such a way as to ensure the bath
will have a minimum of 6 inches of clearance for air circulation (see Figure 2).
Screw the brackets securely to the floor.
5.6.2
Installation of the Bath
Slide the back casters of the bath completely into the bracket caster slots (see
Figure 2). Lock the front casters of the bath. Check that the bath is securely
locked into the brackets by gently pushing on the bath. Proceed to fill the bath
with the applicable bath fluid after reading the entire User's Guide. If you have
any questions concerning installation of the tipping prevention brackets, please
contact a Hart Authorized Service Center (see Section 1.3).
19
5 Installation
Power