Multi-Product Calibrator
Detailed Specifications
23
AC Power (45 Hz to 65 Hz) Specification Summary, PF=1
Current Range
3.3 to
8.999 mA
9 to
32.999 mA
33 to
89.99 mA
90 to 329.99 mA
Voltage
Range
Absolute Uncertainty, tcal
±
5
°
C,
±
(% of watts output)
[1]
33 to 329.999 mV
0.13
0.09
0.13
0.09
90 days
330 mV to 1020 V
0.11
0.07
0.11
0.07
33 to 329.999 mV
0.14
0.10
0.14
0.10
1 year
330 mV to 1020 V
0.12
0.08
0.12
0.08
Current Range
[2]
0.33 to
0.8999 A
0.9 to
2.1999 A
2.2 to
4.4999 A
4.5 to
20.5 A
Voltage
Range
Absolute Uncertainty, tcal
±
5
°
C,
±
(% of watts output)
[1]
33 to 329.999 mV
0.12
0.10
0.12
0.10
90 days
330 mV to 1020 V
0.10
0.08
0.11
0.09
33 to 329.999 mV
0.13
0.11
0.13
0.11
1 year
330 mV to 1020 V
0.11
0.09
0.12
0.10
[1]
To determine ac power uncertainty with more precision, see the individual “DC Voltage Specifications” and “DC Current
Specifications” and “Calculating Power Uncertainty.”
[2]
Add 0.02 % unless a settling time of 30 seconds is allowed for output currents >10 A or for currents on the highest two current
ranges within 30 seconds of an output current >10 A.
Power and Dual Output Limit Specifications
Frequency
Voltages
(NORMAL)
Currents
Voltages
(AUX)
Power Factor
(PF)
dc
0 to
±
1020 V
0 to
±
20.5 A
0 to
±
7 V
⎯
10 to 45 Hz
33 mV to 32.9999 V
3.3 mA to 2.99999 A
10 mV to 5 V
0 to 1
45 to 65 Hz
33 mV to 1020 V
3.3 mA to 20.5 A
10 mV to 5 V
0 to 1
65 to 500 Hz
330 mV to 1020 V
33 mA to 2.99999 A
100 mV to 5 V
0 to 1
65 to 500 Hz
3.3 to 1020 V
33 mA to 20.5 A
100 mV to 5 V
0 to 1
500 Hz to 1 kHz
330 mV to 1020 V
33 mA to 20.5 A
100 mV to 5 V
0 to 1
1 to 5 kHz
3.3 to 500 V
33 mA to 2.99999 A
100 mV to 5 V
0 to 1
5 to 10 kHz
3.3 to 250 V
33 to 329.99 mA
1 to 5 V
0 to 1
10 to 30 kHz
3.3 V to 250 V
33 mA to 329.99 mA
1 V to 3.29999 V
0 to 1
Notes
The range of voltages and currents shown in “DC Voltage Specifications,” “DC Current Specifications,” “AC Voltage (Sine Wave)
Specifications,” and “AC Current (Sine Wave) Specifications” are available in the power and dual output modes (except minimum
current for ac power is 0.33 mA). However, only those limits shown in this table are specified. See “Calculating Power Uncertainty” to
determine the uncertainty at these points.
The phase adjustment range for dual ac outputs is 0
°
to
±
179.99
°
. The phase resolution for dual ac outputs is 0.01 degree.