Multi-Product Calibrator
Detailed Specifications
13
Noise
Range
Bandwidth 0.1 Hz to 10 Hz p-p
±
(ppm
floor)
Bandwidth 10 Hz to 10 kHz rms
0 to 329.9999 mV
0 + 1
μ
V 6
μ
V
0 to 3.299999 V
0 + 10
μ
V 60
μ
V
0 to 32.99999 V
0 + 100
μ
V 600
μ
V
30 to 329.9999 V
10 + 1 mV
20 mV
100 to 1020.000 V
10 + 5 mV
20 mV
Auxiliary Output (dual output mode only)
[1]
0 to 329.9999 mV
0 + 5
μ
V 20
μ
V
0.33 to 3.299999 V
0 + 20
μ
V 200
μ
V
3.3 to 7 V
0 + 100
μ
V 1000
μ
V
[1]
Two channels of dc voltage output are provided.
DC Current
Absolute Uncertainty, tcal
±
5
°
C
±
(ppm of
μ
A)
Range
90 days
1 year
Resolution
Max Compliance
Voltage V
Max Inductive
Load mH
0 to 329.999
μ
A
120 + 0.02
150 + 0.02
1 nA
10
0 to 3.29999 mA
80 + 0.05
100 + 0.05
0.01
μ
A
10
0 to 32.9999 mA
80 + 0.25
100 + 0.25
0.1
μ
A
7
0 to 329.999 mA
80 + 2.5
100 + 2.5
1
μ
A
7
0 to 1.09999 A
160 + 40
200 + 40
10
μ
A
6
1.1 to 2.99999 A
300 + 40
380 + 40
10
μ
A
6
0 to 10.9999 A
(20 A Range)
380 + 500
500 + 500
100
μ
A
4
11 to 20.5 A
[1]
800 + 750
[2]
1000 + 750
[2]
100
μ
A
4
400
[1]
Duty Cycle: Currents <11 A may be provided continuously. For currents >11 A, see Figure 1-4. The current may be provided 60-
T-I minutes any 60 minute period where T is the temperature in
°
C (room temperature is about 23
°
C) and I is the output current
in amperes. For example, 17 A, at 23
°
C could be provided for 60-17-23 = 20 minutes each hour. When the 5522A is outputting
currents between 5 and 11 amps for long periods, the internal self-heating reduces the duty cycle. Under those conditions, the
allowable "on" time indicated by the formula and Figure B is achieved only after the 5522A is outputting currents <5 A for the "off"
period first.
[2]
Floor specification is 1500
μ
A within 30 seconds of selecting operate. For operating times >30 seconds, the floor specification is
750
μ
A.