life of the calibrator and enhance its performance by observing the following rules:
•
The area around the air filter must be at least 3 inches from nearby walls or rack
enclosures.
•
The exhaust perforations on the sides of the calibrator must be clear of obstructions.
•
The air entering the instrument must be at room temperature: make sure the exhaust
air from another instrument is not directed into the fan inlet.
•
Clean the air filter every 30 days or more frequently if the calibrator is operated in a
dusty environment. (See Chapter 7, “Maintenance” for instructions on cleaning the
air filter.)
Where to Go from Here
To locate specific information in the Operators Manual (provided as a pdf file on the
CD-ROM) refer to the following list:
•
Unpacking and setup: Chapter 2, “Prepare for Operation”
•
Installation and rack mounting: Chapter 2, “Prepare for Operation,” and the rack
mount kit instruction sheet
•
AC line power and interface cabling: Chapter 2, “Prepare for Operation”
•
Controls, indicators, and displays: Chapter 3, “Features”
•
Front panel operation: Chapter 4, “Front Panel Operation”
•
Cabling to a UUT (Unit Under Test): Chapter 4, “Front Panel Operation”
•
Remote operation (IEEE-488 or serial): Chapter 5, “Remote Operation”
•
Calibrating an Oscilloscope: Chapters 9, or 10, “SC-600 Oscilloscope Calibration
Option”, or “SC-1100 Oscilloscope Calibration Option”
•
Calibrating Power Quality Equipment: Chapter 11, “PQ Option”
•
Accessories to the 5522A Calibrator: Chapter 8, “Accessories”
•
Performance Specifications: Chapter 1, “Introduction and Specifications”
Summary of Contents for 5522A
Page 4: ......
Page 8: ...5522A Getting Started iv...
Page 10: ...5522A Getting Started vi...
Page 42: ...5522A Getting Started 32...