Page 46
FW-F010-P-M_v0404_01_EN.docx
Input
Flowmeter
Type P
Coil/sine wave (minimum 20mVpp or 80mVpp - sensitivity selectable), NPN/PNP, open
collector, reed-switch, Namur, active pulse signals.
Frequency Minimum 0 Hz - maximum 7 kHz for flow rate.
Maximum frequency depends on signal type and internal low-pass filter.
E.g. Reed switch with low-pass filter: max. frequency 120 Hz.
K-Factor 0.000010 - 9,999,999 with variable decimal position.
Low-pass filter Available for all pulse signals.
Type ZF
coil sensitivity 10mVpp.
Type ZG
coil sensitivity 5mVpp.
Operational
Operator functions
Displayed functions
•
flow rate.
•
measuring and time unit.
•
piegraph 0-100%.
Flow rate
Digits
5 1/2 digits.
Unit
mL, L, m3, Gallons, KG, Ton, lb, bl, cf, RND, ft3, scf, Nm3, Nl, igal - no units.
Time unit
sec - min - hr - day
Decimals
0 - 5.