background image

17 

 

6.7.

 

Calibration 

Calibration can be performed to improve the accuracy of concentration measurements but is 
not required for most proteins and peptides, as the degree of labelling is consistent for the 
majority of samples. Calibration will not affect the size measurement as the rate of diffusion, 
directly measured by the instrument, has a well defined physical relationship as described 
by the Stokes-Einstein equation.  

For outlying proteins which are particularly lysine rich or poor, we recommend performing a 
one-point calibration run using a known concentration of your protein of interest. Full details 
on how to perform a calibration can be found on our website 

www.fluidic.com

  

 

 

Instrument priming

 

The instrument will automatically run a prime sequence to ensure that fluidic lines are fully 
wetted whenever it is required. This will occur whether the instrument was turned on or off 
during the idle period. 

A prime sequence will run when there have been no tests run or the unit has been powered 
down for 8 hours or more, or after a new cartridge has been connected. 

The prime sequence will begin automatically when the first measurement run is started. 

The prime takes around 10 or 20 minutes, depending on how much time has elapsed since 
the last measurement run on the instrument. 

The normal measurement run will begin straight after the prime sequence. 

For the result given immediately after a prime, check the trace graph. If this looks normal, 
the result can be used. If the trace is unusual in any way the result may have been affected 
by the prime 

 in this instance repeat the test. See 

www.fluidic.com

 for tips on interpreting 

traces. 

 

 

Summary of Contents for Fluidity One

Page 1: ...IFU 004 User Guide Issue 4...

Page 2: ...nt Damage caused by accident or misuse Damage caused by disaster Corrosion due to use of improper sample The information contained in this manual is subject to change without notice Fluidic Analytics...

Page 3: ...rtation and handling 10 5 4 Preparing the instrument 10 5 5 Operating conditions 10 5 6 Connections 10 Operation 12 6 1 Manual checks 12 6 2 Turning on and off 12 6 3 Pipetting onto chips 13 6 4 Loadi...

Page 4: ...2...

Page 5: ...ocks Business Centre Cherry Hinton Road Cambridge CB1 8DH UK This manual provides a guide for the installation and upkeep of the Fluidity One by trained service personnel and users All users must fami...

Page 6: ...product has been found to contain SVHC 1 2 dimethoxyethane ethylene glycol dimethyl ether EGDME CAS number 110 71 4 The product is compliant with the following directives provided it is used with the...

Page 7: ...provided the unit is not opened The Fluidity One contains a fan This does not present a danger to users provided the unit is not opened 2 2 Noise Testing was carried out on the instrument to measure...

Page 8: ...provided with the reagent cartridge for hazards associated with the reagent cartridge Reference MSDS001 2 5 Manual handling The Fluidity One weighs 16 kg and should be carried with the appropriate lev...

Page 9: ...y analysis study of protein conformation aggregation or oligomerization and analysis of protein interactions 3 2 Chips Samples are loaded sequentially into the Fluidity One using disposable microfluid...

Page 10: ...experiment is running Reagent cartridges are disposable and should not be re filled by users For full instructions on changing a reagent cartridge see section 6 5 New cartridge Blanking caps cover ea...

Page 11: ...f the instrument are shown below 5 2 Instrument location The Fluidity One is designed for use on a flat laboratory bench Prior to placing the instrument ensure the work surface on which the instrument...

Page 12: ...secure flat and dry work surface Remove all outer packaging from the instrument 5 5 Operating conditions The instrument should be operated at room temperature 15 25 C Take care to position it away fr...

Page 13: ...oad results C Rear USB 2 0 B port for use by Fluidic Analytics engineers during assessment only D Ethernet connection can be used to connect the instrument to a network where WiFi is not available E M...

Page 14: ...5 Ensure there is no obstruction in the chip insert slot 6 2 Turning on and off Connect the power supply to the instrument then to the mains To turn the instrument on press the power button on the ba...

Page 15: ...tion of the sample onto the chip If overflow occurs discard the chip and perform the test again 5 L is the recommended volume 6 4 Loading chips Chips are loaded by sliding them into the slot at the fr...

Page 16: ...f the connector firmly and lifting it off 4 Carefully lift out and remove the old cartridge from the instrument 5 Unpackage the new cartridge and remove the coloured blanking covers on the ports Place...

Page 17: ...e filled 6 6 Starting the measurement Following chip insertion select the appropriate flow rate by choosing the expected size range for your measurement as seen in the following screenshot To start th...

Page 18: ...eriment using the Edit Tag button this includes pre prepared tags like Concentration and Buffer or you can create your own Tags can be edited prior to during and after the run Any tag data will be inc...

Page 19: ...ill automatically run a prime sequence to ensure that fluidic lines are fully wetted whenever it is required This will occur whether the instrument was turned on or off during the idle period A prime...

Page 20: ...en or access advanced results and view the measurement trace Press Finished to return to the home screen Press New experiment to return to the size range selection and start a new measurement Press Ad...

Page 21: ...side Touch on this box to expand and view more detail on the errors See www fluidic com for help with interpreting error codes If you return to the home screen you can access your previously generate...

Page 22: ...stored After this results should be downloaded onto a USB stick and deleted from the instrument to make room for more 11 1 How to download results Download results by following these steps 1 Insert U...

Page 23: ...ent users is changing the cartridge see section 6 5 All other maintenance adjustments or fault repairs must be performed by Fluidic Analytics staff or an authorized agent only To report a fault reques...

Page 24: ...22...

Page 25: ...44 0 1223 560 432 www fluidic com help fluidic com Document reference IFU004 Issue 4...

Reviews: