15
10.
Type in the desired cartridge identifier
–
today’s date is recommended to record the
date of the change, but this can be any value you choose. P
ress “update cartridge
ID” to save.
11.
The low cartridge % will still be displayed until an experiment is run or the
instrument is restarted.
Reagent cartridges are disposable and should not be re-filled.
6.6.
Starting the measurement
Following chip insertion, select the appropriate flow rate by choosing the expected size
range for your measurement as seen in the following screenshot.
To start the measurement, touch Start and then confirm by touching Start Experiment on
the following screen.
Do not attempt to remove the chip from the instrument while an experiment is running and
once you have pressed start.
Experiment times for each size range are approximately as follows:
•
0.5
–
5 nm
–
8 minutes
•
1
–
8 nm
–
10 minutes
•
2
–
20 nm
–
15 minutes