ST51A/ST75A/ST75AV HART Device Specific Commands
In the HART protocol all commands defined as Manufacturer Specific, or Device Specific start at command 128. Use the device specific
commands to setup and configure the ST51A/ST75A/ST75AV instrument via HART.
below summarizes the instrument’s HART Device Specific command set and the data associated with each command.
Table 14 – HART Device Specific Commands
Command 137: Read Totalizer And Rollover Values
Byte
Format
Description
Request Data Bytes
None
Response Data Bytes
0–3
Float
Totalizer
4–7
Integer
Rollover
Response Codes:
See
, for response code list.
Command 138: Read Totalizer State
Byte
Format
Description
Request Data Bytes
None
Response Data Bytes
0
Unsigned-8
Totalizer State: 0 = OFF; 1 = ON
Response Codes:
See
, for response code list.
Command 139: Write Totalizer State
Byte
Format
Description
Request Data Bytes
0
Unsigned-8
Totalizer State: 0 = OFF; 1 = ON
Response Data Bytes
0
Unsigned-8
Totalizer State: 0 = OFF; 1 = ON
Response Codes:
See
, for response code list.
Command 140: Read Device Information
Byte
Format
Description
Request Data Bytes
None
Response Data Bytes
0–9
Bits
Device CO
10–19
Bits
Device Serial Number
19–22
Bits
Device Software Version
Response Codes:
See
, for response code list.
Command 141: Write Totalizer Reset
Byte
Format
Description
Request Data Bytes
0–1
Unsigned-16
Reset Key = 0xABCD
Response Data Bytes
0–1
Unsigned-16
Reset Key = 0xABCD
10–19
Bits
Device Serial Number
19–22
Bits
Device Software Version
Response Codes:
See
, for response code list.
Command 145: Read Customer Engineering Units
Byte
Format
Description
Request Data Bytes
None
Response Data Bytes
0
Unsigned-8
Units Code for Flow
1
Unsigned-8
Units Code for Temperature
2
Unsigned-8
Units Code for Totalizer
Response Codes:
See
, for response code list.
Summary of Contents for ST51A
Page 1: ...Doc No 06EN003426 Rev C...
Page 83: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 75...
Page 84: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 76 Fluid Components International LLC...
Page 85: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 77...
Page 86: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 78 Fluid Components International LLC...
Page 87: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 79...
Page 88: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 80 Fluid Components International LLC...
Page 89: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 81...
Page 90: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 82 Fluid Components International LLC...
Page 91: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 83...
Page 92: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 84 Fluid Components International LLC...
Page 93: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 85...
Page 94: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 86 Fluid Components International LLC...
Page 95: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 87...
Page 96: ...APPENDIX A DRAWINGS ST51A ST75A ST75AV Mass Flow Meters 88 Fluid Components International LLC...
Page 97: ...ST51A ST75A ST75AV Mass Flow Meters APPENDIX A DRAWINGS Fluid Components International LLC 89...