Flowonix Prometra Patient Therapy Controller Manual Download Page 24

Patient Therapy Controller (PTC

®

) Guide, For use with Prometra

®

 Programmable Infusion Systems 

 

24 

Appendix

 

Technical Specifications and EMC Tables 

The Patient Therapy Controller is intended for use in the electromagnetic environment specified in the following EMC tables. The user of the device 
should assure that it is such an environment.  
 

Essential Performance of the Clinician Programmer and PTC 

Medical Equipment Performance Criteria - unacceptable operating conditions / responses are: 
When charging in standby mode the Clinician Programmer or PTC does not go into operating mode. 
When in wireless link mode, the Clinician Programmer or PTC does not stop communication with the pump. It is acceptable for the Clinician 
Programmer or PTC to be temporarily unable to communicate with the pump during the period it is exposed to sources of electromagnetic 
interference, provided that the programmer resumes communication once the exposure period has ended. Electromagnetic interference has been 
classified as a minor risk and does not have an impact on safety. 
 

Specification of RF Frequency Bands 

Bluetooth is the only RF Frequency used by the Clinician Programmer and PTC for transmission and reception. Bluetooth frequency is 2400 MHz. 
 

Table 1 – Guidance and Manufacture’s declaration – Electromagnetic Emissions –  
for all ME Equipment and ME Systems

 

The Patient Therapy Controller is intended for use in the electromagnetic 

environment specified below. The customer and/or the user of this device should 

assure that it is operated in such an environment.

 

RF 

emissions 

CISPR 11

 

Group 1, Class B

 

These components use 

RF energy only for their 

internal functions.

 

Therefore, RF emissions 

are  very low and are not 

likely to cause any 

  

 

 

 

Harmonic emissions

 

 

IEC 61000-3-2

 

Class A

 

These components are 

suitable for use in all 

establishments, including 

domestic establishments 

and those directly 

connected to the public 

low- voltage power 

supply network that 

supplies buildings used 

for domestic purposes.

 

Voltage 

fluctuations/flicker 

emissions

 

 

IEC 61000-3-3

 

Complies

 

 

 

 

Summary of Contents for Prometra Patient Therapy Controller

Page 1: ...Prometra Patient Therapy Controller PTC Guide For use with Prometra Programmable Infusion Systems MR Unsafe Caution Federal Law USA restricts this device to sale by or on the order of a physician...

Page 2: ...ep dry Non ionizing electromagnetic radiation Electrical Safety Type BF Applied Part Waste must be controlled according to local regulation and collection schemes for disposal of Batteries Latex Free...

Page 3: ...evice must accept any interference received including interference that may cause undesired operation IMPORTANT Changes or modifications to this product not authorized by Flowonix Medical Inc could vo...

Page 4: ...view 8 PTC Orientation 9 Warnings 10 Precautions 11 How to Initiate a Bolus using the PTC 11 Troubleshooting 15 PTC Battery Messages 15 PTC Wireless Communication Troubleshooting 17 Touchscreen Troubl...

Page 5: ...be delivered by your pump Drug Bolus Duration the length of time that it will take to deliver your bolus Allotted Boluses per day The maximum number of boluses allowed to be delivered within 24 hours...

Page 6: ...ble side effects or your current treatment regimen is not decreasing your symptoms The use of the Patient Therapy Controller may help you better manage your drug delivery therapy from your Programmabl...

Page 7: ...s not a complete list of possible side effects If you notice other effects not listed above contact your doctor Radio Signals Radio signals sent by your Patient Therapy Controller could interfere with...

Page 8: ...ength of time that you must wait between bolus deliveries If you have any questions concerning the use of your Patient Therapy Controller call Flowonix Medical s Technical Solutions Department at 855...

Page 9: ...rogrammable Pump follow these steps 1 Horizontally position the Patient Therapy Controller up to 2 inches 5cm above the pump The top of the Patient Therapy Controller should be parallel to the top of...

Page 10: ...cal devices has not been established Schedule regular refill visits to avoid reservoir depletion and possible discomfort Contact your physician if you experience changes in your symptoms Potential com...

Page 11: ...Disassembling the Patient Therapy Controller may damage it or cause it to malfunction Certain electronic equipment may cause interference with the Patient Therapy Controller Interference may also occ...

Page 12: ...To charge plug in your included Micro USB Charger into the USB port located at the bottom of the Patient Therapy Controller Then connect it to an AC outlet When charging the Patient Therapy Controlle...

Page 13: ...ear a click when it is pressed The amount of charge will be displayed at the top of the screen Note Charge will need to be above 5 to deliver bolus 3 Once your Patient Therapy Controller is setup it w...

Page 14: ...until the display reads that bolus delivery was successful and success tone is heard During this process the device will display Programming in progress 5 If you have not positioned the Patient Thera...

Page 15: ...he maximum number of boluses allowed has been reached the countdown timer may display an extended lockout period For example Drug Bolus Duration 30 minutes Lockout Duration 2 hours Allotted Boluses pe...

Page 16: ...depending on the charge level of your device This message reminds the user that the device cannot be operated at the same time it is charging The patient does not need to wait for a full charge to de...

Page 17: ...The Patient Therapy Controller has received a full charge of the battery To use the Patient Therapy Controller unplug from the charger At 10 battery the charged icon will turn red The Patient Therapy...

Page 18: ...cm If the Patient Therapy Controller is not positioned properly over the pump it may not be able to communicate with the pump and the bolus delivery may fail See the Patient Therapy Controller Orient...

Page 19: ...not currently set to this pump Contact physician Press OK to continue Your Programmable Pump is not setup with your Patient Therapy Controller Contact physician to setup your Patient Therapy Controll...

Page 20: ...Technical Solutions Department identify the issue with the pump Below is an example of what an error message would look like In all cases contact your physician or Technical Solutions Department for...

Page 21: ...leave it as is Then dry the device with a towel and contact your physician Caution If device becomes wet do not use as there is a possible electrical shock hazard Do not store the device in hot or co...

Page 22: ...lly time out after 1 minute of no operation Voltage Rating 5 VDC Regulatory i Degree of electrical insulation Type BF body floating ii Protection against water ingress Ordinary equipment IP22 iii Mode...

Page 23: ...ix or a refund of the net price paid Wear and tear from normal use or defects resulting from misuse of this product are not covered by this limited warranty To the extent allowable by applicable law t...

Page 24: ...e exposure period has ended Electromagnetic interference has been classified as a minor risk and does not have an impact on safety Specification of RF Frequency Bands Bluetooth is the only RF Frequenc...

Page 25: ...for input output lines Mains power quality should be that of a typical home healthcare environment Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to...

Page 26: ...onment Guidance Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3 V 0 15 MHz 80 MHz 6 V in ISM bands and amateur radio bands between 0 15 MHz and 80 MHz 80 AM at 1 kHz 10 V m 80 MHz 2 7 GHz 80 AM...

Page 27: ...tween this date and product use the user should contact Flowonix Medical Inc to see if additional product information is available Infumorph is a registered trademark of West Ward Pharmaceuticals Corp...

Reviews: