Patient Therapy Controller (PTC
®
) Guide, For use with Prometra
®
Programmable Infusion Systems
22
Patient Therapy Controller Specifications
Height
4.60 inches (11.69 cm)
Width
3.20 inches (8.13 cm)
Depth
1.20 inches (3.05 cm)
Weight
Approximately 190 grams
Telemetry Range
Approximately 2 inches (5.08 cm) above the pump. Interference
may occur in the vicinity of equipment marked with the
following symbol:
Power Source
Rechargeable Lithium ion battery.
Use Environment*
Store between 32° F and 104° F
Use the device between 68° F and 104° F
Relative Humidity: 60+/-15%
Atmospheric Pressure: 860 hPa to 1,060 hPa
Rated to operate to an altitude of 2000 m
Timeout
The programmer will automatically time out after 1 minute of
no operation.
Voltage Rating
5 VDC
Regulatory
i.
Degree of electrical insulation: Type BF (body floating)
ii.
Protection against water ingress: Ordinary equipment
(IP22)
iii.
Mode of operation: Continuous
iv.
Power source: Model
v.
Protection against electric shock: Class II power supply
used
vi.
Anesthetic Warning: This equipment not suitable for
use in the presence of a FLAMMABLE ANESTHETIC
MIXED WITH AIR or OXYGEN or NITROUS OXIDE.
vii.
Means of Operator Protection (MOOP) for Power
Supply and Means of Patient Protection (MOPP) for
Patient Therapy Controller.
viii.
Standards used for Patient Therapy Controller:
MEDICAL GENERAL MEDICAL EQUIPMENT
AS TO ELECTRICAL SHOCK, FIRE AND MECHANICAL
HAZARDS ONLY IN ACCORDANCE WITH ANSI/AAMI
ES60601-1 (2005), CAN/CSA-C22.2 No. 60601-1
*See Appendix 1 for additional information regarding Electromagnetic Compatibility.