SERVICE
MANUAL
10
EASYVAC VACUUM REGULATOR MS003 February 2010
Maintenance operations
2.3.4
REPLACEMENT
OF THE VACUUM GAUGE
If the periodic checks on the accuracy value of the vacuum gauge indicated in the operating manual of the
EASYVAC device shows this value falls outside the tolerance levels or in the case of accidental damage to the
vacuum gauges, they must be replaced, as specified in the Table of illustrations on page 4.
WARNING: REPLACE THE VACUUM GAUGES USING ONLY ORIGINAL SPARE PARTS WITH THE
SAME FULL-SCALE VALUE AS THE ONES BEING REPLACED.
The replacement methods are shown with the assistance of the following illustrations:
Use a TORX screwdriver of suitable dimensions to
unscrew the two screws on the rear lid.
After removing the lid, take the vacuum gauge out
of the body of the regulator, pulling it outwards,
ensuring it is moved at a right angle to the body.