EASYVAC VACUUM REGULATOR              MS003           February 2010 

 

 

Contents 

Warning ..............................................................  

Your local dealer and service center ..................  

Table of illustrations ............................................   

 

Drawings ............................................................  

List of spare parts ...............................................  

Maintenance operations .....................................   

 

 

  Warning 

WARNING: DANGER INDICATION 

Attention: Important indication

 

General information 

 

Read this manual carefully before using the 

EASYVAC vacuum regulator.

 

 

THIS DOCUMENT MUST BE USED TOGETHER 

WITH AND IN ACCORDANCE WITH THE 

OPERATOR MANUAL SUPPLIED WITH THE 

PRODUCT.

 

 

AFTER UNPACKING AND CONNECTION, 

PERFORM THE FUNCTIONAL TEST DESCRIBED 

IN THE CHAPTER "PERIODIC CHECKS" OF 

OPERATING MANUAL MO002 ON THE DEVICE.

 

 

USE OF INSTALLATION PROCEDURES NOT 

DESCRIBED IN THE OPERATOR MANUAL 

MAY REDUCE THE DEVICE’S SAFETY LEVEL.

 

 

BEFORE EACH USE, ENSURE THAT THE 

EASYVAC VACUUM REGULATOR IS 

DIRECTLY OR INDIRECTLY CONNECTED TO A 

COLLECTION SYSTEM FITTED WITH AN 

OVERFLOW VALVE AND ANTIBACTERIAL 

FILTER AND THAT THE CONNECTOR HOSES 

USED CONFORM WITH THE STANDARD EN 

ISO 10079-3.

 

 

The company accepts no responsibility for the 

non-observance of these service instructions, 

the use of non-original spare parts and/or 

repairs by unauthorised persons.

 

Connections

 

 

DO NOT ALLOW OIL AND GREASE TO COME 

INTO CONTACT WITH THE DEVICE.

 

 

The vacuum supply connector is positioned on 

the rear of the device and can be identified by 

referring to the drawing on page 5 of 

Instructions Manual MO002 (pos. 7).

 

Operation 

 

THE EASYVAC VACUUM REGULATOR MUST 

ALWAYS BE USED WITH CARE AND ONLY BY 

PERSONNEL WHO ARE AWARE OF THE 

CONSEQUENCES OF THE ONGOING 

THERAPY.

 

 

Disconnect the vacuum supply when the unit 

is not in use.

 

Cleaning and disinfection

 

DO NOT USE SOLVENTS OR ABRASIVE PRODUCTS 

FOR CLEANING: THESE CAN SERIOUSLY DAMAGE 

THE SURFACES OF THE EQUIPMENT AND THE 

PLASTIC PARTS. 

 

DO NOT IMMERSE THE UNIT IN DISINFECTANT; 

 

DO NOT PLACE THE UNIT IN AN AUTOCLAVE; 

 

DO NOT USE INFLAMMABLE PRODUCTS. 

IF THE EQUIPMENT IS ACCIDENTALLY CONTAMINATED 
BY SUCTIONED LIQUIDS, IT MUST BE ISOLATED AND 
THE NECESSARY CLEANING AND DISINFECTION 
OPERATIONS CARRIED OUT BY QUALIFIED 
PERSONNEL. 
IF NECESSARY, STERILISE THE EASYVAC VACUUM 
REGULATOR WITH A VALIDATE ETHYLENE OXIDE 
STERILIZATION CYCLE (E.G. 14% EO 12 H. / 50 KPA / 
37 °C).

 

Service

 

 

All the modifications and repairs must only be 
performed by personnel authorized by 
FLOW METER S.p.A., or by hospital 
technicians approved by the same company. 

 

Original spare parts must always be used for 
the maintenance operations. 

 

Check the EASYVAC vacuum regulator every 
three years in accordance with the chapter 
“Periodic checks” of Instructions Manual 
MO002. 

 

Due to periodic updates, the construction of 
the appliance may vary. On this point, 
FLOW METER guarantees the availability of 
spare parts for a minimum period of 5 years 
from the date of manufacture. 

 

Any modification to the device should only be 
performed if approved by the manufacturer  
and in accordance with the procedures 
provided.

 

 

The device has been designed and 
manufactured to meet the safety requirements 
of the following standards: 

 

EN ISO 10079-3

 

 

Summary of Contents for EASYVAC

Page 1: ...SERVICE MANUAL SERVICE MANUAL EASYVAC VACUUM REGULATOR PUBLICATION CODE MS003 EDITION no 6 DATE February 2010...

Page 2: ...YVAC VACUUM REGULATOR MUST ALWAYS BE USED WITH CARE AND ONLY BY PERSONNEL WHO ARE AWARE OF THE CONSEQUENCES OF THE ONGOING THERAPY Disconnect the vacuum supply when the unit is not in use Cleaning and...

Page 3: ...n of FLOW METER S p A FLOW METER S p A appliances are subjected to periodic revisions to maintain and improve the production standards functioning and increase their reliability For this reason the co...

Page 4: ...SERVICE MANUAL 4 EASYVAC VACUUM REGULATOR MS003 February 2010 Table of illustrations Drawing no TITLE PAGE 1 1 Spares for EASYVAC vacuum regulator 5...

Page 5: ...SERVICE MANUAL 5 EASYVAC VACUUM REGULATOR MS003 February 2010 Drawing no 1 1...

Page 6: ...00 I O button switch complete with gaskets 3a 000010192 Vacuum gauge 1000 mbar for EASYVAC regulator complete with fixing support gaskets and protection cover 3b 000010193 Vacuum gauge 600 mbar for EA...

Page 7: ...ry 6 months 2 3 3 REPLACEMENT OF THE LID WITH A MEMBRANE every 24 months 2 3 4 REPLACEMENT OF THE VACUUM GAUGE in the case of accuracy outside the tolerance levels or breakage 2 3 MAINTENANCE OPERATIO...

Page 8: ...urned with the indication 0 towards the front The rear slot of the button must fit into the relative coupling inside the body of the vacuum regulator itself Keeping the green button pressed down push...

Page 9: ...mpened with ethyl alcohol diluted to 60 in water Reattach the lid complete with the regulating knob and the membrane onto the body of the vacuum regulator tightening it manually for as long as resista...

Page 10: ...he vacuum gauges they must be replaced as specified in the Table of illustrations on page 4 WARNING REPLACE THE VACUUM GAUGES USING ONLY ORIGINAL SPARE PARTS WITH THE SAME FULL SCALE VALUE AS THE ONES...

Page 11: ...m gauge Carefully insert the vacuum gauge unit back into the body aligning the vacuum gauge holder block guides with those on the body of the EASYVAC vacuum regulator pushing it right in Be careful to...