SERVICE
MANUAL
7
EASYVAC VACUUM REGULATOR MS003 February 2010
Maintenance operations
2.
MAINTENANCE
2.1
INTRODUCTION
This section provides instructions for proper maintenance of the device, in order to guarantee correct functioning
and appropriate conditions of safety.
2.2
FREQUENCY OF MAINTENANCE
It is recommended that the operations specified in chapter 2.3 MAINTENANCE OPERATIONS be performed with
the following frequency:
2.3.1 REPLACEMENT OF THE I-0 BUTTON
in the case of breakage or loss of seal
2.3.2 CLEANING OF THE MEMBRANE AND BODY
every 6 months
2.3.3 REPLACEMENT OF THE LID WITH A MEMBRANE
every 24 months
2.3.4 REPLACEMENT OF THE VACUUM GAUGE
in the case of accuracy outside the tolerance levels or breakage
2.3
MAINTENANCE
OPERATIONS
2.3.1 REPLACEMENT OF THE I-0 BUTTON
If the I-0 closing tap shows signs of leakage or is damaged, the entire kit (2), shown in the drawing specified in the
Table of illustrations on page 4, must be replaced.
The replacement methods are provided with the assistance of the following illustrations:
Holding the EASYVAC vacuum regulator firmly,
pull out the red button, until it is completely free of
its quick coupling, and remove it completely.
Using a screwdriver, push the green button
outwards, until it is completely removed from the
vacuum regulator.