EASYVAC VACUUM REGULATOR              MS003           February 2010 

 

 

Maintenance operations 

2.  

MAINTENANCE 

 

 2.1 

INTRODUCTION 

This section provides instructions for proper maintenance of the device, in order to guarantee correct functioning 
and appropriate conditions of safety. 

 

2.2 

FREQUENCY OF MAINTENANCE 

It is recommended that the operations specified in chapter 2.3 MAINTENANCE OPERATIONS be performed with 
the following frequency: 

2.3.1 REPLACEMENT OF THE I-0 BUTTON 

in the case of breakage or loss of seal 

2.3.2 CLEANING OF THE MEMBRANE AND BODY 

every 6 months 

2.3.3 REPLACEMENT OF THE LID WITH A MEMBRANE 

every 24 months 

2.3.4 REPLACEMENT OF THE VACUUM GAUGE 

in the case of accuracy outside the tolerance levels or breakage 

 2.3 

MAINTENANCE 

OPERATIONS 

 

2.3.1  REPLACEMENT OF THE I-0 BUTTON 

If the I-0 closing tap shows signs of leakage or is damaged, the entire kit (2), shown in the drawing specified in the 
Table of illustrations on page 4, must be replaced.  
The replacement methods are provided with the assistance of the following illustrations: 

 

 

Holding the EASYVAC vacuum regulator firmly, 
pull out the red button, until it is completely free of 
its quick coupling, and remove it completely. 

Using a screwdriver, push the green button 
outwards, until it is completely removed from the 
vacuum regulator. 

 
 

Summary of Contents for EASYVAC

Page 1: ...SERVICE MANUAL SERVICE MANUAL EASYVAC VACUUM REGULATOR PUBLICATION CODE MS003 EDITION no 6 DATE February 2010...

Page 2: ...YVAC VACUUM REGULATOR MUST ALWAYS BE USED WITH CARE AND ONLY BY PERSONNEL WHO ARE AWARE OF THE CONSEQUENCES OF THE ONGOING THERAPY Disconnect the vacuum supply when the unit is not in use Cleaning and...

Page 3: ...n of FLOW METER S p A FLOW METER S p A appliances are subjected to periodic revisions to maintain and improve the production standards functioning and increase their reliability For this reason the co...

Page 4: ...SERVICE MANUAL 4 EASYVAC VACUUM REGULATOR MS003 February 2010 Table of illustrations Drawing no TITLE PAGE 1 1 Spares for EASYVAC vacuum regulator 5...

Page 5: ...SERVICE MANUAL 5 EASYVAC VACUUM REGULATOR MS003 February 2010 Drawing no 1 1...

Page 6: ...00 I O button switch complete with gaskets 3a 000010192 Vacuum gauge 1000 mbar for EASYVAC regulator complete with fixing support gaskets and protection cover 3b 000010193 Vacuum gauge 600 mbar for EA...

Page 7: ...ry 6 months 2 3 3 REPLACEMENT OF THE LID WITH A MEMBRANE every 24 months 2 3 4 REPLACEMENT OF THE VACUUM GAUGE in the case of accuracy outside the tolerance levels or breakage 2 3 MAINTENANCE OPERATIO...

Page 8: ...urned with the indication 0 towards the front The rear slot of the button must fit into the relative coupling inside the body of the vacuum regulator itself Keeping the green button pressed down push...

Page 9: ...mpened with ethyl alcohol diluted to 60 in water Reattach the lid complete with the regulating knob and the membrane onto the body of the vacuum regulator tightening it manually for as long as resista...

Page 10: ...he vacuum gauges they must be replaced as specified in the Table of illustrations on page 4 WARNING REPLACE THE VACUUM GAUGES USING ONLY ORIGINAL SPARE PARTS WITH THE SAME FULL SCALE VALUE AS THE ONES...

Page 11: ...m gauge Carefully insert the vacuum gauge unit back into the body aligning the vacuum gauge holder block guides with those on the body of the EASYVAC vacuum regulator pushing it right in Be careful to...