35
Help with therapy problems
Problem
Cause
Solution
The Therapy device
does not start
or
the
pressure is much too
high and the therapy
device does not
react
Pressure gauge hose
not connected to the
therapy device
Connect pressure gauge hose
and therapy device
Water in pressure
gauge hose
Dry pressure gauge hose
Pressure gauge hose
kinked or blocked
Check pressure gauge hose
is not blocked, replace
breathing hose as necessary,
contact your supplier
Pressure gauge hose is
closed with stopper
Remove stopper
Hose smells
unpleasant
Breathing hose smells
due to manufacturing
Allow the breathing hose to
air
The breathing hose is
contaminated with
germs
Replace breathing hose
Therapy device does
not regulate itself
correctly
The internal cross-
section of the pressure
gauge hose has been
reduced, e.g. through
crushing
Replace the hose, contact
your supplier
Pressure gauge hose is
too short (does not
extend into mask
connection)
Change the breathing hose
or have the supplier adjust it
Water on the
heating plate
Possible hairline crack
in the beaker
Replace the beaker, contact
your supplier
Internal noise
development
______
Contact your supplier or
service engineer
Whatever the problem: If you are unable to resolve it yourself, please contact
your supplier or the doctor in charge of the treatment.
Summary of Contents for Vigaro
Page 1: ...Instruction manual FLO Vigaro Next Generation English Rev 1 0 July 2018 Vigaro from REV 4 15...
Page 3: ...Instruction manual FLO Vigaro Next Generation English July 2018...
Page 4: ......
Page 61: ...61 Notice...
Page 62: ...62 Notice...
Page 64: ...Art Nr 40000245...