![Flo-tech mc4000 User Manual Download Page 4](http://html1.mh-extra.com/html/flo-tech/mc4000/mc4000_user-manual_2306497004.webp)
Product Unpacking And Inspection
PRODUCT UNPACKING AND INSPECTION
Upon receipt of the product, perform the following unpacking and inspection procedures .
OTEE:
N
If damage to the shipping container is evident upon receipt, request the carrier to be present when the product
is unpacked .
1 . Carefully open the shipping package, follow any instructions that may be marked on the exterior . Remove all cushioning
material surrounding the product and carefully lift the product from the package .
2 . Save the package and all packing material for possible use in reshipment or storage .
3 . Visually inspect the product and applicable accessories for any physical damage such as scratches, loose or broken parts or
any other sign of damage that may have occurred during shipment .
OTEE:
N
If damage is found, request an inspection by the carrier’s agent within 48 hours of delivery and file a claim with the
carrier . A claim for equipment damage in transit is the sole responsibility of the purchaser .
INTRODUCTION
The MC4000 has inputs for two pressure sensors, one temperature sensor, one flow sensor, and one active pickup for RPM
measurements . The MC4000 allows simultaneous connection of four sensors with the measurements displayed in four
individual LCD windows using preselected process units according to US or DIN norms . The display supports English and one
of the following languages selected at the time of order: Spanish, German, Italian, or French .
An internal rechargeable battery powers the MC4000 . A green LED indicates a completely charged battery . A fully charged
battery permits four hours of operation with two pressure sensors connected . With the charger connected, the battery
charges while also allowing use of the instrument . An optional automobile style power cable is available for operation from a
car battery .
When in
Datalogger
mode, the MC4000 stores data in an internal 2 .5 MB memory space . Each stored measurement contains
the day and date from the instruments real time clock .
The MC4CON software utility transfers stored measurements via a USB data port to a Windows PC . The MC4CON program
permits bi-directional communication not only for uploading recorded data to a PC, but also for downloading commands
from the PC to the MC4000 .
All parameters can also be set using the buttons on the instrument’s front panel . Pressure units are programmed in psi or bar .
The flow and rpm measurements are scaled using programmable constants . Additionally, three point flow calibration for the
turbine sensors is available . When activated, the pressure tare function allows offset adjustments and the sets the display to
zero at non-zero pressure inputs .
Hydraulic horsepower is automatically calculated from the measured pressure and flow . Horsepower displays as either HP or
kW depending on the selected units .