MC4CON Software
RPM
RPM parameter settings
Scale rpm
Multiplication constant for rpm scaling
Dscale rpm
Division constant for rpm scaling
New-Tab
Three point linearizing of flow sensors
Example
Tab F1 INP
First frequency point
0 Hz
Tab Q1 OUT
First displayed flow reading
0 LPM
Tab F2 INP
Second frequency point
166 Hz
Tab Q2 OUT
Second displayed flow reading
38 LPM
Tab F3 INP
Third frequency point
630 Hz
Tab Q3 OUT
Third displayed flow reading .
150 LPM
Units
Flow units are calculated in US or European standard units
EU
Default units are lpm, bar, ° C
US
Default units are gpm, psi, ° F
Language
Selection of English or default (the local country language)
English
Default
Backlight
Backlight control
Light off
Backlight is turned off, no instrument backlight
Light on
Backlight is turned on the display and illuminates for 30 seconds when any buttons are pressed
File
Menu MC4000, upper tree switches
New Device
Default setting
File Read
Reads all the menu parameters from a file with the file format
filename
.
set
from the MC4000
File Save
Stores all the menu parameters in a file with the format
filename
.
set
to the MC4000
Device
Menu MC4000, lower three switches
Menu MC4000 read
Reads the stored MC4000 parameters into the MC4CON software
Menu MC4000 store
Reads the stored MC4CON parameters into the MC4000
Timer MC4000 write
Transfer time and date values from the PC into MC4000