6.2.
Removing the lens interface
Lenses are fragile elements.
Take the greatest care at the time of their handling, particularly on
the level of the lenses.
Lenses can be heavy.
Some lenses weight several hundred grams. Be careful not to be
surprised by its weight.
The lens interface of the camera, which also holds the filter wheel
mechanism, can be removed so a custom lens interface can be installed.
Turn gently the optic counter-clockwise while placing a hand below. It may
be necessary to do some turns.
Place caps on the lens and store it.
Never force on lens.
Check that optics is clean before assembling it in the camera. See
section 7.1 for the maintenance of optics
Unscrew the six screws located around the M80 thread using the
appropriate 2.5mm BTR key.