![Flann Microwave SD5902 Operation And Maintenance Manual Download Page 6](http://html1.mh-extra.com/html/flann-microwave/sd5902/sd5902_operation-and-maintenance-manual_2302683006.webp)
Flann Microwave
SD5902 Waveguide Switch Driver
SD5902
Doc No ER15 Version 1.3
June 2020
6
Specifications
System Performance
Max Switching Time (2 Channel)
180ms* (Speed Mode), 475ms* (Precision Mode – worst case)
(3 Channel)
250ms* (Speed Mode), 500ms* (Precision Mode – worst case)
*Note that the indicated switching times are measured from the initial command to the motor stopping, and
do not include the rotor settling time. Consult the Flann Microwave sales team for more details.
Power Requirements
This is a safety class 1 product (provided with a protective earth terminal). A non-interruptible safety
earth ground must be provided from the main power source to the Waveguide Switch Driver power
input terminals and power cord. Whenever the safety earth ground has been impaired, the
instrument must be made inoperative and secured against any unintended operation. If this
instrument is to be energised via an autotransformer (NOT RECOMMENDED) for voltage
reduction, make sure that the common terminal is connected to the earth pole of the power source.
Line Voltage
90 – 264 V AC
Line Frequency
47-63 Hz
Power Consumption
50W Max (Motor Running), 25W Max (Motor Stationary)
Supply Fuse
20mm T2.5A Slow Blow (On rear panel)
Electrical Specifications
GPIB Connector
24-Way, compatible with Amphenol 57 Series (IEEE488 Standard)
Switch Output Connector
15-Way D-Type Female
Switch Cable Length
5m Max
Motor Steps per Revolution
500
Mechanical Specifications
Operating Orientation
Any
Dimensions (mm)
58H × 483W × 300D mm (2.28H × 19.0W × 11.8D inch)
Weight
2.15 kg (4.74 lb)
Direction of Rotation
Unidirectional: Counter-clockwise when viewed from the top
Environmental Specifications
Operating Temperature Range
0 °C to +40 °C
Storage Temperature Range
-20 °C to +60 °C
Humidity 90%,
No
condensation