![Flaem RF6 Basic 2 Operating Instructions Manual Download Page 12](http://html1.mh-extra.com/html/flaem/rf6-basic-2/rf6-basic-2_operating-instructions-manual_2302006012.webp)
10
TECHNICAL SPECIFICATIONS
RF6 Basic
2
Nebulizer
In compliance with 93/42/EEC Directive
Minimum fill volume: 2 ml
Maximum fill volume: 8 ml
APPLIED PARTS
The applied parts of type BF are:
patient accessories (1, 2, 3, 4, 5, 6, 7, 8, 9)
Operating conditions:
Temperature: min 10°C; max 40°C
Air humidity: min 10%; max 95%
Atmospheric pressure: min 69KPa; max 106KPa
Storage conditions:
Temperature: min -25°C; max 70°C
Air humidity: min 10%; max 95%
Atmospheric pressure: min 69KPa; max 106KPa
FLAEM
compres-
sor range
Ope-
rating
pressure
(bar)
MMAD
(μm)
(2)
BREATHABLE FRACTION
< 5 μm (FPF) %
(2)
DELIVERY ml/min
(1)
(approx)
standard
mode
high
speed
mode
standard
mode
high speed
mode
standard
mode
high speed
mode
F400
0.60
3.8
4.00
62
63
0.25
0.32
F700 -
F1000
0.80
3.6
3.6
66
67
0.32
0.42
(1)
Data shown is in accordance with Flaem Nuova internal procedure I29-P07.5.
(2)
In vitro characterization
carried out at the Inamed Research GmbH Co. KG on behalf of TÜV Rheinland LGA Products GmbH – Germany in
compliance with the new European Standard for aerosol therapy units, Standard EN 13544-1. Further details are
available upon request.
Summary of Contents for RF6 Basic 2
Page 34: ...32 SOFTTOUCH SoftTouch 13 3 3 4 5 12 RF6 Basic2 RF6 Basic2 5 2 high speed...
Page 44: ...42 SOFTTOUCH SoftTouch 13 3 3 4 5 10 RF6 Basic2 2 RF6 Basic2 5 ml 2...
Page 49: ...47 SOFTTOUCH SoftTouch 4 10 10 10...
Page 52: ...MEMO MEMORANDUM NOTIZEN NOTAS...
Page 53: ...MEMO MEMORANDUM NOTIZEN NOTAS...
Page 54: ...MEMO MEMORANDUM NOTIZEN NOTAS...