TROUBLE-SHOOTING
Switch off the device before any procedure and unplug the power cable from the socket.
PROBLEM
CAUSE
SOLUTION
The device does not work
The power cable has not
been correctly inserted into
the power socket
Correctly insert the power cable
in sockets.
The device does not
nebulize or
nebulizes insufficiently.
The medication has not been
inserted in the nebulizer
Pour the right amount of
medication in the nebulizer
The nebulizer has not been
properly fitted.
Disassemble the nebulizer and
reassemble it correctly as shown
in the connection diagram on
the cover.
The nebulizer nozzle is
clogged
Disassemble the nebulizer,
remove the nozzle and perform
cleaning operations.
Failure to remove medication
deposits from the nebulizer
affects its efficiency and
operation.
Strictly comply with the
instructions
contained in the “CLEANING,
SANITISATION, DISINFECTION,
STERILISATION” chapter
The air pipe is not correctly
connected to the device
Ensure the device’s accessories
are properly connected to
the air intake (see connection
diagram on the cover).
The air pipe is bent, damaged
or twisted
Unwind the pipe and ensure it is
not crushed or punctured. If
necessary replace it.
The air filter is dirty
Replace the filter
The apparatus is noisier
than usual
The Filter is not correctly
inserted in its housing
Fully and properly insert the
Filter in its housing
If after verifying the above mentioned conditions the device should not operate properly,
contact your trusted retailer or the nearest authorized FLAEM service center.