8
9
Mod.: P0611EM F1000
Max Pressure:
2.6 ± 0.4 bar
Compressor air output:
10 l/min approx
Noise level (at 1 m):
54 dB (A) approx
Operation:
Continuous use
Dimensions (L)x(P)x(H):
16x16x10 cm
Weight:
1.45 Kg
Operating conditions:
Temperature:
min 10°C; max 40°C
Air humidity:
min 10%; max 95%
Atmospheric pressure:
min 69KPa; max 106KPa
Storage conditions:
Temperature:
min -25°C; max 70°C
Air humidity:
min 10%; max 95%
Atmospheric pressure:
min 69KPa; max 106KPa
APPLIED PARTS
Type BF applied parts include: (C2, C3, C4)
RF7 DUAL SPEED PLUS Nebulizer
Medication minimum capacity: 2 ml
Medication maximum capacity: 8 ml
Operating pressure (with neb.): 0.90 bar
TECHNICAL DATA
Voltage:
230V~ 50Hz 140VA 115V ~ 60Hz 220V ~ 60Hz 100V ~ 50/60Hz
Safety certifications:
Speed selector C1.4
in pos Max
Speed selector C1.4
in pos Min with valve
(1)
Delivery:
0.53 ml/min approx.
0.23 ml/min approx.
(2)
MMAD:
2.91 μm
2.76 μm
(2)
Breathable fractions < 5 μm (FPF):
76%
78.5%
(1)
data detected according to Flaem I29-P07.5 internal procedure
(2)
In vitro characterization certified by TÜV Rheinland Product Safety GmbH - Germany in compliance with the new
European Standard for aerosol therapy units, Standard EN 13544-1, ANNEX CC. Further details are available on request.
DEVICE DISPOSAL
In compliance with the Directive 2012/19/EC, the symbol printed on the device
shows that the device to be disposed of is considered waste and must therefore be
an item of ''differentiated collection''. Consequently, the user must take it (or have it
taken) to the designated collection sites provided by the local authorities, or turn it in to
the dealer when purchasing an equivalent new device. Differentiated waste collection and
the subsequent treatment, recycling and disposal procedures promote the production of
devices made with recycled materials and limit the negative effects on the environment
and on health caused by potential improper waste management. The unlawful disposal
of the product by the user could result in administrative fines as provided by the laws
transposing Directive 2012/19/EC of the European member state or of the country in
which the product is disposed of.