SOFTTOUCH FACE MASKS
Soft
biocompatible
material
Dispersion
Limiting Vent
SoftTouch
masks have an outer edge made of
soft biocompatible material
that ensures excel-
lent adherence to the face, and is also equipped
with an
innovative Dispersion Limiting Device
.
These distinctive elements that distinguish it allow
greater sedimentation of medication in the patient
and also
limit dispersion
.
During the inspiratory phase,
the
tab
, acting as a
Dispersion
Limiting Device
, bends inwards
towards the mask.
During the expiratory phase, the
tab
, acting as a
Dispersion Lim-
iting Device
, bends outwards
from the mask.
CLEANING SANITISATION DISINFECTION
Switch off the device before any cleaning procedure and unplug the power cable from
the socket.
DEVICE AND TUBING EXTERIOR
Use only a damp cloth with antibacterial soap (non-abrasive and with no solvents
of any sort).
ACCESSORIES
Open the nebuliser by turning the upper part (C1.3) anticlockwise, remove the noz-
zle (C1.2) from the upper part (C1.3). Disassembly by pressing as shown by the 2
arrows in the “Connection diagram” in point C1.2.
Then proceed according to the following instructions.
SANITISATION
Before and after each use, sanitise the nebuliser cup and the accessories, choosing
one of the methods described below.
method A:
Sanitise accessories
C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C4
under potable
hot water (approximately 40 °C) with a gentle, non abrasive dish detergent.
method B:
Sanitise accessories
C1.1-C1.2-C1.3-C1.4-C2-C3-C4
in the dishwasher
with a hot cycle.
method C:
Sanitise accessories
C1.1-C1.2-C1.3-C1.4-C2-C3-C4
by immersing them
in a solution of 50% water and 50% white vinegar, then rinse thoroughly under pota-
ble hot water (approximately 40 °C).
If you want to also perform the cleaning for DISINFECTION, jump to the DISINFEC-
TION paragraph.
After having sanitised the accessories, shake them vigorously and place them on a
paper towel. Alternatively, dry them with a jet of hot air (for example, a hair dryer).
DISINFECTION
After sanitising the nebuliser cup and the accessories, disinfect them choosing one
of the methods described below.
method A:
Accessories
C1.1-C1.2-C1.3-C1.4-C2-C3-C4
can be disinfected.
The disinfectant must be an electrolytic chloroxidizer (active principle: sodium hy-
pochlorite) specific for disinfecting, which is available in any pharmacy.
Implementation:
- Fill a container big enough to hold all of the parts to disinfect with a solution of
potable water and disinfectant, according to the proportions indicated on the
packaging of the disinfectant.
4
5