2  |   Important Notes

Digital Drybath

Important Notes

General Practice

Safety 

During operation, maintenance and repair of this instrument, 
the following basic safety notes should be observed. In case 
of failure to follow these instructions, the warnings or notes 
indicated herein, the basic protection provided by the 
instrument, its safety criteria of design and manufacture, and 
its predicted use range would be impaired. If the equipment is 
used in a manner not specified by the manufacturer, the 
protection provided by the equipment may be impaired.

CAUTION: 

Important information is contained in 

any item and should be carefully read. Failure to 
observe the instruction would result in damage or 
abnormal function of the instrument.

WARNING: 

The warning message requires 

extremely careful operation of a certain step. 
Failure to observe the instruction may result in 
serious personal injury.

CAUTION: 

This instrument is for indoor use.

The ground connection

To avoid the electric shock, the input power 
line must be reliably grounded. The instrument 
is equipped with the three-pin plug that has 
the third pin (the pin connects the ground), 
therefore, the plug should be used with the 
grounded power socket only. This is a kind of 
safety device. If the plug cannot be plugged 
into the power socket, please ask the 
electrician to install a correct power socket, so 
as to make the grounded plug to work for 
safety.
Fisher Scientific is not responsible for any 
injury as a result of the user’s failure to 
observe the following requirements.

Servicing and Replacement of Parts

The operator should not take the instrument 
apart without permission. Only qualified 
professionals are allowed to replace elements 
or adjust components inside the instrument. 
Replacement of components while unit is 
connected to power and/or turned on is 
prohibited.

Observe voltage/power requirements

Before the instrument is connected to the AC 
power source, the voltage of the power 
source should be the same with the required 
voltage of the instrument (a deviation of ±10% 
is allowed). The rated load of the power 
socket should not be lower than the 
requirements of the instrument.

The power cord

The instrument should use the power line cord 
attached to it. If the power line is destroyed, it 
must be replaced but not be repaired. The 
replacement should be carried out with the 
power line of the same type and same 
specification (reach out to technical service 
team for further information). No items should 
be put on the power cord when the 
instrument is in operation.

Connecting and disconnecting to/from 
power source

The user should hold the plug to remove from 
power source. When connecting the plug, 
user should make sure it has been fully 
plugged in; when removing the plug, do not 
pull the power line forcefully. Do not position 
the equipment so that it is difficult to operate 
the disconnecting device (plug).

Placement of the instrument

This instrument should be fixed in a low 
RH (

≤ 

80%) and low dust place away from 

water source (e.g. sink or water pipeline) and 
the room should be well ventilated, and free of 
corrosive gas or interference of strong 
magnetic field. The instrument should not be 
placed in a wet or dusty location.

Summary of Contents for 88860021

Page 1: ...Digital Drybath Operational Manual 0000853 Revision C October 2021...

Page 2: ...ic be held liable for any damages direct or incidental arising out of or related to the use of this manual No part of this manual may be reproduced or transmitted in any form or by any means including...

Page 3: ...ions 6 Basic Parameters 6 Preparation Work 8 Structure Diagram 8 Display Panel Diagram 8 Instructions for Buttons 8 Operation Guidance 9 Examination Before Startup 9 Startup 9 Setting Temperature and...

Page 4: ...Warranty 15 Frequently Asked Questions 16...

Page 5: ...Vac FS Drybath Stdrd 1 blck 100 120 V 88860022 N A 100 120 Vac FS Drybath Stdrd 2 blck 100 120 V 88860023 N A 100 120 Vac FS Drybath Stdrd 4 blck 100 120 V 88860024 15387928 200 240 Vac FS Drybath Std...

Page 6: ...erve the following requirements Servicing and Replacement of Parts The operator should not take the instrument apart without permission Only qualified professionals are allowed to replace elements or...

Page 7: ...ecting the mains plug to a standard socket outlet Always place the product in such a way that it is easy for the operator to disconnect the product from the mains supply Explanation of symbols Attenti...

Page 8: ...Safety Data Sheets SDS or regulations applicable to your country and region This product has materials and components that may be recycled or reused according to local guidelines and regulations Remo...

Page 9: ...ration and serum coagulation etc The characteristics of the product are as follows The digital display for control of the temperature Heat blocks help avoid possible contamination while containing the...

Page 10: ...strument please make sure the working condition meets the above requirements Especially that the power line is reliably grounded Table 2 100 120 V NA Catalog Number 88860021 88860022 88860023 Model Na...

Page 11: ...ature Uniformity 1 0 C 37 C Temperature Accuracy 0 5 C 37 C Temperature Fluctuation 0 5 C Temperature Rise Time 20 min rise from 30 C to 130 C 25 min rise from 30 C to 130 C Size L W H 288 200 100 mm...

Page 12: ...ift blocks Digital Display Warning Indicator Light Heating Indicator Light Setting or adjusting button Press the button to display present setting temperature and press the button again to display pre...

Page 13: ...ress Stop 4 Set up Temperature a Press the SET CAL button The instrument is in the Temperature Setpoint mode with the display window showing the previous temperature set value and the left digit flash...

Page 14: ...e Display and independent measurements And check if the temperature gap is 0 2 C or not 5 If less than 0 2 C no calibration should be made 6 If more than 0 2 C calibrate the Temperature by these steps...

Page 15: ...ith the wall of the holes so as to have good heat conduction If the surface of the instrument is polluted it can be cleaned with a slightly damp soft cloth Generally unit falls under protective system...

Page 16: ...1 Display window is not turned on after switching on power supply Power supply fails to be connected Check power supply and connect it Fuse burned out Replace fuse Damage of the switch Replace the swi...

Page 17: ...860107 For 17 mm 18 mm test tubes 12 124 x 76 x 39 4 9 x 3 0 x 1 5 88860108 For 20 mm test tubes 8 124 x 76 x 39 4 9 x 3 0 x 1 5 88860109 For 25 mm test tubes 6 124 x 76 x 39 4 9 x 3 0 x 1 5 88860110...

Page 18: ...24 x 76 x 113 4 9 x 3 0 x 4 55 88860118 For 15 ml conical bottom test tubes 15 124 x 76 x 113 4 9 x 3 0 x 4 55 88860119 For 50 ml conical bottom test tubes 4 124 x 76 x 113 4 9 x 3 0 x 4 55 88860120 F...

Page 19: ...these operation instructions and maintenance this product is warranted for 24 months against defective materials or workmanship The 24 month warranty period begins from the delivery date of this prod...

Page 20: ...oesn t hold the set temperature It will cool to ambient temp Q Would unit return to ambient temp A The unit will return to ambient temp Scenario 2 Suppose the unit is running a set program and there i...

Page 21: ...allen finden Sie unter fisher com WEEERoHS Conformit WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2012 19 EU in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici W...

Page 22: ...43 0 800 20 88 40 at fishersci com Netherlands 31 0 20 4887 70 00 nl fishersci com Belgium 32 0 56 260 260 be fishersci com Norway 47 22 95 59 59 fishersci no Denmark 45 70 27 99 20 fishersci dk Port...

Page 23: ...Road Asheville NC 28804 United States Find out more at fishersci com 2021 Fisher Scientific Inc All rights reserved All trademarks are the property of Fisher Scientific and its subsidiaries unless ot...

Reviews: