| Start-up
22
|
50145736
Fisherbrand
Chapter 5
1. Loosen the two screws [1] in the bottom panel, then remove the entire bottom panel by lifting
it out of its two embossings.
Figure 5-5.
Bottom Panel Removed
2. Grab the two retaining springs [1] at their tabs and pull them downwards out of the
embossings, then pull off the lateral air baffle.
Figure 5-6.
Removing the Lateral Air Baffle
3. For the lateral air baffles to install correctly, the two retaining springs [1] must be facing
upwards. Place the lateral air baffle on the lower embossings and tilt it upwards against the
side wall of the work space.
4. Clamp the two retaining springs [1] into the upper embossings.
5. Replace the bottom panel into the embossings and secure it by fastening the two screws at [1].
Risk of damage to sensor!
To prevent accidental damage, be sure to avoid collision with the sensor when
installing or uninstalling panels in the oven‘s workspace.
Summary of Contents for Fisher Sei 100L Gravty
Page 1: ......
Page 5: ...Fisherbrand 50145736 3 Chapter 15 Spare Parts and Accessories 51 Chapter 16 Device Log 53 ...
Page 6: ...4 50145736 Fisherbrand ...
Page 8: ...2 50145736 Fisherbrand ...
Page 14: ... Safety Notes 6 50145736 Fisherbrand Chapter 1 ...
Page 34: ... Start up 26 50145736 Fisherbrand Chapter 5 ...
Page 40: ... Handling and Control 32 50145736 Fisherbrand Chapter 7 ...
Page 42: ... Shut down 34 50145736 Fisherbrand Chapter 8 ...
Page 46: ... Cleaning and Disinfection 38 50145736 Fisherbrand Chapter 9 ...
Page 52: ... Disposal 44 50145736 Fisherbrand Chapter 11 ...
Page 54: ... Error Codes 46 50145736 Fisherbrand Chapter 12 ...
Page 60: ... Spare Parts and Accessories 52 50145736 Fisherbrand Chapter 14 ...
Page 62: ... Device Log 54 50145736 Fisherbrand Chapter 15 ...