Fisher Scientific 146E Quick Start Manual Download Page 5

TROUBLE SHOOTING and REPAIR 

Symptom 

Possible Cause 

Possible Solution to Condition 

Incubator controller isn’t 
functioning, no lights etc.  

Power switch isn’t ON or no power to 
unit.  

Place Power ON/ OFF switch inside chamber to 
ON.  Check power outlet by plugging a known 
good device into it. 

Erratic temperature display 
readings above or below set 
point. 

Frost buildup. Look for frost inside 
chamber around evaporator.   

Defrost per instructions.  If possible operate in 
50% Cooling Mode to reduce defrosting cycles. 

Temperature uniformity in 
chamber above specified limit. 

Frost buildup or airflow blocked by 
samples. 

Defrost per instruction.  If airflow blocked by 
sample containers, remove some to allow top to 
bottom airflow. 

Temperature readings too 
high or too low compared to 
thermometer. 

Controller may need calibrated.  

Refer to CALIBRATION section of this manual 

Control goes into power on 
reset sequence while 
operating. 

Inadequate line power. Test Line 
voltage of outlet to Incubator while unit 
is operating. 

If line voltage is < 106 VAC while unit is running, 
then a higher capacity AC line is required. 

Compressor not operating 
while COOL LED is lit. 

Compressor stalled and thermal 
overloads opens. Test line voltage of 
outlet while unit is operating. 

If line voltage is < 106 VAC while unit is running, 
then a higher capacity AC line is required. 

Temperature readings go far 
below setpoint and trigger 
alarm at 15°C. 

Open heater or heater control relay 
SSR1. 

Unplug unit from power then refer to replacing 
HTR1. Check heater with OHM meter, about 29 
ohms (97-995E 60 ohms). 

Temperature readings go far 
above setpoint and triggers 
alarm at 40°C or above. 

Defective cooling relay, compressor or 
shorted heater SSR1.  Check if 
compressor is running, if so touch 
evaporator in chamber, it should be 
cold. 

If compressor isn’t running while Cool led is on 
then replace K2 relay. If compressor is running 
and evaporator isn’t cold then cooling system 
needs repair.  If evaporator is cold then SSR1 
may be shorted. 

Controller display works but 
heater, compressor and 
chamber fan don’t. 

Safety relay K1 not closing. 

K1 relay may need replaced. 

Display EEE/99.9 & Alarm 1 

Open sensor or sensor connection. 

Check sensor connection at controller. Check 
sensor with ohm meter, 100 ohms at 0°C and 
110 ohms at 25°C. 

Display EEE/99.8 & Alarm 1 

Shorted sensor or connection. 

Check sensor connection at controller. 

Display EEE/temp & Alarm 1 

Incubator needs defrosted. 

See erratic temperature display readings. 

1. Temp Limit   = Temp (+/-3°C or more from setpoint) &  (> 40°C or <15°C), safety relay opens.

 

 

Replacement Parts 

 
Description  

 

Spare Part Number 

 

 Description 

  Spare 

Part 

Number 

TC1 

Controller 

  SPN102214 

  S1 

Switch 

  SPN83373 

Sensor, RTD  

 SPN88613 

  HTR1 

Heater 

(115V) 

 SPN102079 

SSR1 

Relay 

 

  SPN88616 

  HTR1 

Heater 

(230V) 

 SPN102080 

K1 

Safety 

Relay 

  SPN102260 

  FAN2 

(115V) 

  SPN30580 

K2 

Comp. 

Relay  SPN102259 

  FAN2 

(230V) 

  SPN39854 

T1 

Transformer 

  SPN88678 

  THM1 

Thermostat 

 SPN88612 

Compressor (97-990E)   

SPN105501 

 
Technical Assistance:   

1-800-926-0505 

Customer Service: 

 

1-800-766-7000 

 
Fisher Scientific  
2000 Park Lane 
Pittsburgh, PA 15275 USA 
 
Instruction Manual 
Part # 95525 Rev. J 

 

Summary of Contents for 146E

Page 1: ...ment must be earth grounded for safe operation Maximum ratings of accessories that may be plugged into convenience outlet Load current 2 Amps 115VAC Leakage Current 250 micro Amps IMPORTANT When operating at 100 cooling the unit should be defrosted periodically especially during times of high humidity The customer will be responsible for any repair work done caused by evaporator ice build up See D...

Page 2: ...erence CONTROLS Power ON OFF Switch Located inside chamber area is used to turn power 1 ON or 0 OFF to entire unit Keys Located on front panel MENU Used to access menus SET Used to display setting INCREASE setting DECREASE setting DISPLAY and INDICATORS DISPLAY On front panel Displays chamber and set point temperatures in C and prompts from menu INDICATORS Front panel DELAY Lights when delay start...

Page 3: ...es at temperatures 10 C or higher above ambient 35 C to 60 C It provides the greatest power savings yet still maintain the advertised stability No defrosting required Pro Menu Protection Menu Used to select the Protection mode of operation that determines when the alarm is activated and if the control setpoint can be changed Range 0 0 to 0 2 0 0 Mode Normal mode that allows the user to change setp...

Page 4: ...eometric center reading 5 Enter into the Calibration Mode as described in Accessing Menu 6 Press and hold SET key using the or keys set the calibration parameter to the value determined in step 4 7 Allow unit to stabilize for 1 hour and check that controller value and geometric center value agree 2 C Repeat calibration if necessary Replacing Temp Controller 1 Disconnect incubator from power source...

Page 5: ...y SSR1 Unplug unit from power then refer to replacing HTR1 Check heater with OHM meter about 29 ohms 97 995E 60 ohms Temperature readings go far above setpoint and triggers alarm at 40 C or above Defective cooling relay compressor or shorted heater SSR1 Check if compressor is running if so touch evaporator in chamber it should be cold If compressor isn t running while Cool led is on then replace K...

Page 6: ...Schematic Diagram ...

Reviews: