Fisher Scientific 146E Quick Start Manual Download Page 2

INSTALLATION

 

The Low Temperature Incubator should be installed on a 
level, stable floor.  Install the 4 adjustable feet per 
instructions inside package. Use the adjustable feet to level 
the incubator.  Locate the unit so that area in front of door 
and kick grill is clear.  To prevent heat build up within the 
unit, do not install it where it will be exposed to strong 
sunlight, nor near radiators, furnaces, or other sources of 
heat.  Do not install the incubator where flammable or 
corrosive atmospheres may exist.  The Low Temperature 
Incubator is not to be used out-of-doors or where excessive 
moisture or contamination could cause degrading of 
mechanical or electrical components. 
 
1. Remove all protective tape from doors and shelves. 
2. Clean interior and exterior with warm soap and water.  
Shipping tape residue can be removed using isopropyl 
alcohol. 
3. Uncoil line cord from rear of unit.  Connect the plug to a 
power outlet that complies with the electrical requirements 
specified on the unit's label and with proper safety ground 
connection.  If line cord replacement is necessary, use CSA 
types:  SJ, SJO or SJT, 16 GA, 3 conductor or equivalent.  
 

Door Reversal:

 

1.  Disconnect unit from power source.  On inside of door, 
directly behind controller, open snapped conduit and 
remove controller cable. 
2. Remove screw that secures cable clamp holding 
controller cable, then replace screw. 
3. Remove controller cable from J clamps that run along 
inside chamber side and back walls.  
4. Place 3 J clamps (Provided) at approximately the same 
spacing (on opposite side) as existing J clamps. 
5. Support door with one hand remove screw securing door 
to upper hinge, then lift door off bottom hinge. 
6.  Remove the three screws securing each hinge to 
chassis. 
7. Remove three plug screws covering holes on left side, 
then install them in holes on right side, 
8. Install hinges on left side using screws removed in step 
6.  Position hinge with door screw on bottom. 
9. Position door on hinge then insert screw previously 
removed in step 5. 
10. Attach supplied latching duct to inside on door, in line 
with controller cable toward opposite side on door from 
other duct. 
11.  Route cable using J clamps and latching duct. 
12. Remove screw on left front of enclosure and install 
cable clamp with removed screw. 
13. Adjust size of loop in cable where door is hinged to 
ensure door can fully open and close without interference.  

 
 
CONTROLS

 

Power ON/OFF Switch:

 Located inside chamber area, is 

used to turn power (1) ON or (0) OFF to entire unit. 
 

Keys: 

Located on front panel.  

MENU  Used to access menus.  
SET      Used to display setting. 

 INCREASE setting. 

 DECREASE setting. 

 

DISPLAY and INDICATORS

 

DISPLAY: 

 (On front panel) 

Displays chamber and set point temperatures in 

°

C and 

prompts from menu. 
 

INDICATORS:

  (Front panel) 

DELAY: 

 Lights when delay start timer for compressor is 

timing.   Will begin timing when power is first applied or 
when cooling is first turned off. 

ALARM:

  Indicates that chamber temperature is not within 

limits and the safety relay is opened. 

COOL: 

 Lights when cooling compressor is on. 

HEAT:

  Lights when controller demands heat.   

 
 

GENERAL FUNCTIONAL CHECKS

 

1.Turn power on to unit by placing the Power ON/ OFF 
Switch to I (ON) position. The Controller (mounted on front 
of door) will display S , then, 88.8 (to test display), then 
begins reading actual chamber temperature.   
Note:  When power is first applied, if display indicates an L 
instead of an S, a switch setting on the controller must be 
changed.  Refer to Replacing Temp Controller (TC1). 
2. After 3 seconds, the alarm LED on the controller should 
extinguish and the safety relay should close applying power 
to the heater, fan, compressor relay  and convenience 
outlet. 
3. Press and hold MENU key, after 3 sec. the display will 
indicate CAL.  Release and press MENU again to display 
Pct, press SET, display should indicate 100.  Press and 
release MENU again to indicate Pro, press SET, display 
should indicate 00, press and release MENU to return to 
normal control mode.  Press SET to display control setpoint, 
display should read 25.0.  If the control setpoint or any of 
the other parameters aren't set correctly, use the 

 or 

 

keys while pressing SET to change the setting. 
4. The DELAY LED will be lit for 7 minutes after power is 
first applied, the COOL LED will be out during this delay 
period.   When the DELAY LED goes out, the COOL LED 
will light and the compressor relay will close starting the 
compressor. 
5. Allow unit to stabilize for 1 hour at 25

°

C, to ensure that all 

systems are functioning. 
 
 

OPERATION

 

Perform the procedure under General Functional Checks if 
incubator is new or hasn't been operated for a long time.   
 

Controller operation:

 

Usually the only controller setting that is necessary to 
change is the chamber temperature control setpoint.  Other 
parameters of the controller may be changed by accessing 
them through the MENU key, covered later.     

Changing Chamber Temperature Control Setpoint

 

1.  The chamber temperature is continuously displayed, 
unless a key is pressed. 

Summary of Contents for 146E

Page 1: ...ment must be earth grounded for safe operation Maximum ratings of accessories that may be plugged into convenience outlet Load current 2 Amps 115VAC Leakage Current 250 micro Amps IMPORTANT When operating at 100 cooling the unit should be defrosted periodically especially during times of high humidity The customer will be responsible for any repair work done caused by evaporator ice build up See D...

Page 2: ...erence CONTROLS Power ON OFF Switch Located inside chamber area is used to turn power 1 ON or 0 OFF to entire unit Keys Located on front panel MENU Used to access menus SET Used to display setting INCREASE setting DECREASE setting DISPLAY and INDICATORS DISPLAY On front panel Displays chamber and set point temperatures in C and prompts from menu INDICATORS Front panel DELAY Lights when delay start...

Page 3: ...es at temperatures 10 C or higher above ambient 35 C to 60 C It provides the greatest power savings yet still maintain the advertised stability No defrosting required Pro Menu Protection Menu Used to select the Protection mode of operation that determines when the alarm is activated and if the control setpoint can be changed Range 0 0 to 0 2 0 0 Mode Normal mode that allows the user to change setp...

Page 4: ...eometric center reading 5 Enter into the Calibration Mode as described in Accessing Menu 6 Press and hold SET key using the or keys set the calibration parameter to the value determined in step 4 7 Allow unit to stabilize for 1 hour and check that controller value and geometric center value agree 2 C Repeat calibration if necessary Replacing Temp Controller 1 Disconnect incubator from power source...

Page 5: ...y SSR1 Unplug unit from power then refer to replacing HTR1 Check heater with OHM meter about 29 ohms 97 995E 60 ohms Temperature readings go far above setpoint and triggers alarm at 40 C or above Defective cooling relay compressor or shorted heater SSR1 Check if compressor is running if so touch evaporator in chamber it should be cold If compressor isn t running while Cool led is on then replace K...

Page 6: ...Schematic Diagram ...

Reviews: