Fisher Scientific 146E Quick Start Manual Download Page 3

2. To display the setpoint temperature, press the SET key.  
The setpoint temperature will remain on the display for 1 
sec. after the SET key is released. 
3. To change the setpoint temperature, press and hold SET 
 while pressing the  

 or 

 keys to change the setting. (Note 

that when either 

 or 

 keys are first pressed, the display 

will begin to change slowly at first, then increase change 
rate after a few seconds.) 
4. Select the desired chamber setpoint then release all 
keys.  The controller display will revert back to actual 
chamber temperature after a few seconds.  Allow up to an 
hour for unit to stabilize at the new setpoint temperature. 
 
After the incubator temperature has stabilized, place 
samples into incubator chamber.   For best results, arrange 
samples evenly throughout the chamber.  Liquid samples 
should be covered to prevent evaporation and eventual 
frost build up on evaporator coils, particularly when 
operating below ambient.  Note that frost build-up is normal 
when unit is being operated below 35

°

C with cooling set to 

100% mode.   
 
The Low Temperature Incubator has an internal 2 amp. 
convenience outlet inside chamber to operate devices such 
as: shakers, rotators, photosynthesis lights, recording 
thermometers etc.  

Accessing Controller Menus:

 

The temperature controller has three menu selections that 
can be accessed by pressing and holding the MENU key for 
3 sec.  To access menu parameter, press and hold the SET 
key.  To change the parameter, use the 

 or 

 keys. 

 

CAL Menu

- Calibration Menu is used to change the 

temperature offset value to correct for differences in 
chamber temperature and the displayed temperature.  
Range -5

°

 to 5

°

C.  See Calibration Procedure. 

 

Pct Menu

- (Percent Cooling Menu) Used to select one of 

three cooling modes:  
100% Mode- Compressor runs continuously which provides 
high precision control (+/-.2

°

C Stability) over entire 

temperature range.  Requires manual defrost. 
 
 50% Mode- Compressor runs approximately half the time, 
which conserves power and also reduces the amount of 
frost buildup on the evaporator.  This method greatly 
reduces the frequency of manual defrosting but provides 
less temperature stability (+/- 1.5°C).   
 
00% Mode- Compressor is completely off.  This mode is 
ideal for incubating samples at temperatures 10

°

C or higher 

above ambient (35

°

C to 60

°

C).  It provides the greatest 

power savings yet still maintain the advertised stability.  No 
defrosting required. 

 

Pro Menu

- Protection Menu - Used to select the Protection 

mode of operation that determines when the alarm is 
activated and if the control setpoint can be changed.  
Range 0.0 to 0.2.   
 
0.0 Mode- Normal mode that allows the user to change 
setpoint.  Alarm activates if chamber temperature 
(chamber) and setpoint temperature (setpoint) are as 
follows:  
chamber 

 15.0

°

C and chamber < (setpoint - 3) or 

chamber 

 40.0

°

C and chamber > (s3) 

While ALARM is activated, the safety relay is opened 
interrupting power to the heater, fan, convenience outlet 
and compressor.  When chamber temperature returns to 
acceptable limits, the ALARM will automatically terminate 
and power is reapplied.  In order to allow changing of 
setpoint and eliminate nuisance alarms, an alarm by-pass 
takes effect when the setpoint is changed or when power is 
first applied.   The alarm by-pass works as follows: Present 
chamber temperature is stored as a nominal alarm 
temperature (nominal).  If chamber < setpoint then alarm 
temperature (alarm) = nominal -3.  If chamber > setpoint 
then alarm = n3.  If chamber should go beyond 
alarm or < -13 or > 63 then the ALARM will activate.  This 
alarm by-pass is active until chamber reaches setpoint 
±1

°

C. 

 
0.1 Mode-  Same as 0.0 Mode, except that the setpoint is 
locked and can't be changed. 
 
0.2 Mode  Setpoint  is locked, plus, the ALARM will activate 
and the safety relay will open as follows: 
chamber < (setpoint -3) or chamber > (se3) 
In order to use this mode of protection, modes 0.0 or 0.1 will 
have to be used to reach the setpoint temperature.  Once 
the chamber temperature reaches the setpoint and is 
stable, protection mode 0.2 can be set. 
Note! If the chamber temperature falls outside the setpoint 
by  ±3

°

C, the ALARM will activate and safety relay will 

interrupt power.  This mode can only be used if the door is 
to remain shut and other disturbances don't effect the 
chamber temperature. 
 
As an added safety feature, the incubator has a thermostat 
in the heater compartment that limits the upper chamber 
temperature to 65

°

C.  

 

MAINTENANCE

 

  

WARNING

 

To reduce the risk of electric shock, disconnect from 
power source before servicing.

 

 
Before re-applying power, after maintenance is complete, 
check to ensure that safety ground is intact and making a 
good connection. 

Summary of Contents for 146E

Page 1: ...ment must be earth grounded for safe operation Maximum ratings of accessories that may be plugged into convenience outlet Load current 2 Amps 115VAC Leakage Current 250 micro Amps IMPORTANT When operating at 100 cooling the unit should be defrosted periodically especially during times of high humidity The customer will be responsible for any repair work done caused by evaporator ice build up See D...

Page 2: ...erence CONTROLS Power ON OFF Switch Located inside chamber area is used to turn power 1 ON or 0 OFF to entire unit Keys Located on front panel MENU Used to access menus SET Used to display setting INCREASE setting DECREASE setting DISPLAY and INDICATORS DISPLAY On front panel Displays chamber and set point temperatures in C and prompts from menu INDICATORS Front panel DELAY Lights when delay start...

Page 3: ...es at temperatures 10 C or higher above ambient 35 C to 60 C It provides the greatest power savings yet still maintain the advertised stability No defrosting required Pro Menu Protection Menu Used to select the Protection mode of operation that determines when the alarm is activated and if the control setpoint can be changed Range 0 0 to 0 2 0 0 Mode Normal mode that allows the user to change setp...

Page 4: ...eometric center reading 5 Enter into the Calibration Mode as described in Accessing Menu 6 Press and hold SET key using the or keys set the calibration parameter to the value determined in step 4 7 Allow unit to stabilize for 1 hour and check that controller value and geometric center value agree 2 C Repeat calibration if necessary Replacing Temp Controller 1 Disconnect incubator from power source...

Page 5: ...y SSR1 Unplug unit from power then refer to replacing HTR1 Check heater with OHM meter about 29 ohms 97 995E 60 ohms Temperature readings go far above setpoint and triggers alarm at 40 C or above Defective cooling relay compressor or shorted heater SSR1 Check if compressor is running if so touch evaporator in chamber it should be cold If compressor isn t running while Cool led is on then replace K...

Page 6: ...Schematic Diagram ...

Reviews: