FISCHER MD-1a Technical Reference Manual Download Page 7

Electrical Device Inspection

 

Period: Annually (12 months) 
 

Line Voltage

,

 Ground Integrity:

 

 

Measure line voltage supplied to the device . 

 

Accept: 105-120 VAC

 

 

Confirm reliable ground connection by ensuring line receptacle is 3-prong grounded 
type and ground integrity is intact. 

 

 

 

Accept: Ground OK

 

 

Measure  device  ground  wire  resistance  between  line  plug  ground  contact  and  the 
screw near the Heyco cord grip on the rear panel.   

Accept:  0  to  1.0  ohm       

Nominal  reading  is  .1-.2  ohms. 

IMPORTANT:  Do  not  make  this  measurement 

between line plug ground contact and front or rear panel surfaces; these panels are 
anodized and will not conduct.

 

 

Chassis and Patient Leakage:

 

 

Measure  ground  leakage  between  line  and  chassis  using  an  approved  Electrical 
Safety  Analyser.  Use  screw  head  near  the  Heyco  cord  grip  on  the  rear  panel  for 
chassis connection. 

 

 

 

 

 

Accept:  Less  than  50uA

 

IMPORTANT:  Do  not  make  this  measurement  between  line  and  front  or  rear  panel 
surfaces; these panels are anodized and will not conduct.

  

 

Nominal readings appear below for reference.  

 

Normal Polarity-Normal Ground:   

5 uA 

 

Reverse Polarity-Normal Ground:  

5 uA 

 

Normal Polarity-Open Ground: 

 

10 uA 

 

Reverse Polarity-Open Ground:   

10 uA

 

 

Measure patient lead leakage between line and red and black Output Jacks using an 
approved Electrical Safety Analyser.   

 

 

Accept: Less than 50uA 

 

Nominal readings appear below for reference.  

 

 

Normal Polarity-Normal Ground:   

5 uA 

 

 

Reverse Polarity-Normal Ground:  

5 uA 

 

 

Normal Polarity-Open Ground: 

 

5 uA 

 

Reverse Polarity-Open Ground:   

5 uA 

 

Dielectric Strength Test (OPTIONAL):

 

 

Apply 1250 VAC, 60 Hz between line and chassis ground for 60 seconds using an 
approved Dielectric Withstand Tester, with Cutoff current set to 1.0-2.0 ma. 

 

Apply 1250 VAC, 60 Hz between line and patient leads (red and black jacks) for 60 
seconds using an approved Dielectric Withstand Tester with Cutoff current set to 1.0-
2.0 ma.   

 

 

 

 

 

 

Accept: No Breakdown

 

IMPORTANT: Observe safety precautions when conducting dielectric strength test. 
Do not return to service any device which has failed the withstand test as it may have 
become internally damaged. This test is optional and does not need to be conducted 
annually. 

DEVICE INSPECTION 

Page 6 

Summary of Contents for MD-1a

Page 1: ......

Page 2: ...trols Page 3 Device Specifications Page 4 Device Maintenance Page 4 Device Inspection Page 5 Assembly Parts List Page 8 Schematic Diagram Page 9 MODEL MD 1a Galvanic Unit Technical Reference Manual CO...

Page 3: ...nding although the plug will fit typical grounded receptacles Because of its light weight and small size the MD 1 is very portable When relocating the unit for use however always be sure it is properl...

Page 4: ...witch will be exposed when the power is turned ON INTENSITY KNOB Adjust this knob to set the desired current level to be applied to the patient The actual current is indicated on the DC MILLIAMPERES m...

Page 5: ...Patient Override Control N A Indicator Display Unit Functioning Active lamp Current Level analog DC milliameter Meter ranges 0 10 DCma w 0 2ma resolution 0 50 DCma w 1 0ma resolution Meter Accuracy 20...

Page 6: ...Minimum intensity setting and turn on at a setting of 1 2 on the knob Verify operation of Polarity Reversal switch applies to units made after 1 1 2005 Switch in REV position will reverse polarity at...

Page 7: ...t Nominal readings appear below for reference Normal Polarity Normal Ground 5 uA Reverse Polarity Normal Ground 5 uA Normal Polarity Open Ground 10 uA Reverse Polarity Open Ground 10 uA Measure patien...

Page 8: ...pproximately 50 DCma at this this setting which will require a resistor rated at 2 5 Watts or higher Accept 40 58 DCma Verify meter reading on 0 10 scale With a 1K load resistance set the METER SCALE...

Page 9: ...Resistor 1 00 38 0004 10K 2W 53C3 Pot 1 00 41 0013 100uf 100V Electrolytic 1 00 41 0015 100uf 250V Electrolytic 2 00 41 0033 33uf 25V Tantalum 1 00 45 0015 005uf 50V Ceramic Disc 1 00 45 0006 001uf A...

Page 10: ...Page 9...

Reviews: