FISCHER MD-1a Technical Reference Manual Download Page 4

Description of Model MD-1a panel controls 

DC MILLIAMPERE METER 

This  analog  meter  indicates 

actual

  current  flow,  in  DC  Milliamperes,  through  the  connected  patient 

circuit. 

A reading can only be obtained when a resistive load is applied between output jacks and the 

INTENSITY knob is adjusted. 

 

METER SCALE SWITCH (0-10 and 0-50) 

This switch selects the meter scale to be used for indicating current flow. Read from the TOP range 
(0-10) when the switch is in the "0-10" position, and from the BOTTOM range (0-50) when in the "0-
50"  position.  This  switch  does  not  change  the  level  of  output,  only  the  meter  scale.  Use  the  "0-10" 
scale for a more accurate reading when using currents less than 10 milliamperes. 
 

ACTIVE LAMP 

This  red  LED  will  glow  ONLY  when  the  Galvanic  Unit  detects  both  a  completed  patient  circuit  and 
active current flow. 

This lamp does not indicate a power-on condition, but only current activation.

 

 
POWER SWITCH 

Press the LOWER edge (marked "1") of this rocker-type switch to turn ON the Galvanic Unit. The red 
colored top edge of the switch will be exposed when the power is turned ON. 
 
I

NTENSITY KNOB 

Adjust  this  knob  to  set  the  desired  current  level  to  be  applied  to  the  patient.  The  actual  current  is 
indicated on the "DC MILLIAMPERES" meter when the patient circuit is completed. 

Since current flow 

is  dependent  upon  the  patient  circuit  resistance  (which  varies  with  electrode  contact  area  and  skin 
moisture), only the meter reading can provide an accurate indication of current level - the INTENSITY 
knob position may vary.  
 

POLARITY REVERSAL SWITCH (Applies to units made after 1/1/2005) 

Press the left side for normal (

NOR

) polarity (red jack is positive), or the right side for reverse (

REV

polarity (black jack is positive). Use this switch to reverse polarity instead of swapping the patient cord 
plugs. 

Always return the INTENSITY to “0” before changing polarity. 

 

OUTPUT JACKS 

Two "banana" jacks allow connection of patient cords to the machine. The RED jack (marked "+") is 
always positive with respect to the BLACK (negative) jack. The supplied Dual Patient Cordset uses a 
single,  dual-prong  plug  at  one  end  to  allow  quick  connection  or  reversal  of  electrode  polarity  as 
needed. Always insure that the plug is securely attached to the output jacks during treatment. 
 


 
 
 
 

 
 
 
 
 

 
 
 

 
 
 

 
 
 
 
 
 

 
 
 
 

 

 
 

Page 3 

Summary of Contents for MD-1a

Page 1: ......

Page 2: ...trols Page 3 Device Specifications Page 4 Device Maintenance Page 4 Device Inspection Page 5 Assembly Parts List Page 8 Schematic Diagram Page 9 MODEL MD 1a Galvanic Unit Technical Reference Manual CO...

Page 3: ...nding although the plug will fit typical grounded receptacles Because of its light weight and small size the MD 1 is very portable When relocating the unit for use however always be sure it is properl...

Page 4: ...witch will be exposed when the power is turned ON INTENSITY KNOB Adjust this knob to set the desired current level to be applied to the patient The actual current is indicated on the DC MILLIAMPERES m...

Page 5: ...Patient Override Control N A Indicator Display Unit Functioning Active lamp Current Level analog DC milliameter Meter ranges 0 10 DCma w 0 2ma resolution 0 50 DCma w 1 0ma resolution Meter Accuracy 20...

Page 6: ...Minimum intensity setting and turn on at a setting of 1 2 on the knob Verify operation of Polarity Reversal switch applies to units made after 1 1 2005 Switch in REV position will reverse polarity at...

Page 7: ...t Nominal readings appear below for reference Normal Polarity Normal Ground 5 uA Reverse Polarity Normal Ground 5 uA Normal Polarity Open Ground 10 uA Reverse Polarity Open Ground 10 uA Measure patien...

Page 8: ...pproximately 50 DCma at this this setting which will require a resistor rated at 2 5 Watts or higher Accept 40 58 DCma Verify meter reading on 0 10 scale With a 1K load resistance set the METER SCALE...

Page 9: ...Resistor 1 00 38 0004 10K 2W 53C3 Pot 1 00 41 0013 100uf 100V Electrolytic 1 00 41 0015 100uf 250V Electrolytic 2 00 41 0033 33uf 25V Tantalum 1 00 45 0015 005uf 50V Ceramic Disc 1 00 45 0006 001uf A...

Page 10: ...Page 9...

Reviews: