FISCHER MD-1a Technical Reference Manual Download Page 6

The  factory  recommends  inspecting  this  device  according  to  the  following  schedule. 
Inspections may be performed at other intervals as required by maintenance personnel 
at the installation site, but in no event should the interval between inspections exceed 24 
months. 
 

Visual Device Inspection

 

Period: Annually (12 months) 
 

Line Cord and plug:

  

 

Inspect for insulation wear, broken wires, exposed wires, loose Heyco grip 

 

Inspect for bent or broken plug spades, bent or broken ground connection 

 

Replace line cord upon evidence of wear or breakage 

  

Panel Controls: 

 

Inspect for damaged or worn panel controls. 

 

Rotate Intensity control knob to check for positive stops at Minimum and Maximum. 

 

Tighten Intensity control knob if knob is loose. 

 

Check integrity of Output Jacks and replace if corrosion or cracks are evident. 

 

Verify  operation  of  “Active”  LED  by  applying  a  1K  to  10K  ohm  load  across  Output 
Jacks and rotating Intensity knob. LED should be off at Minimum intensity setting and 
turn on at a setting of 1-2 on the knob. 

 

Verify  operation  of  Polarity  Reversal  switch  (applies  to  units  made  after  1/1/2005). 
Switch in “REV” position will reverse polarity at red & black panel jacks.. 

 

Verify operation of Meter by applying a 1K to 10K ohm load across Output Jacks and 
rotating Intensity knob. Confirm at both meter scales (0-10 and 0-50) with a nominal 
output setting of 10 DCma as indicated on the meter. 

 
Cabinet:

 

 

Inspect case for evidence of cracks or breakage. 

 

Ensure  that  bottom  panel  screws  (4)  are  tight  and  front  and  panels  are  properly 
seated in the housing (only necessary if unit has been opened for access to circuitry). 

 

Visual Electrode Inspection

 

Period: 6 months 
 

Electrodes and Cords: 

 

Inspect patient cords (red and black) for evidence of wear, broken wires, exposed 
wires or corroded banana jacks 

 

Replace patient cords regularly (12-24 months) or if wear is evident. 

 

Inspect dual plug at one end of cords; tighten allen screw if necessary to clamp 
internal wire. 

 

Inspect electrodes (wet pads or Iontophoresis trays) for evidence of breakage or 
corrosion. Replace if necessary. 

 

DEVICE INSPECTION 

Page 5 

Summary of Contents for MD-1a

Page 1: ......

Page 2: ...trols Page 3 Device Specifications Page 4 Device Maintenance Page 4 Device Inspection Page 5 Assembly Parts List Page 8 Schematic Diagram Page 9 MODEL MD 1a Galvanic Unit Technical Reference Manual CO...

Page 3: ...nding although the plug will fit typical grounded receptacles Because of its light weight and small size the MD 1 is very portable When relocating the unit for use however always be sure it is properl...

Page 4: ...witch will be exposed when the power is turned ON INTENSITY KNOB Adjust this knob to set the desired current level to be applied to the patient The actual current is indicated on the DC MILLIAMPERES m...

Page 5: ...Patient Override Control N A Indicator Display Unit Functioning Active lamp Current Level analog DC milliameter Meter ranges 0 10 DCma w 0 2ma resolution 0 50 DCma w 1 0ma resolution Meter Accuracy 20...

Page 6: ...Minimum intensity setting and turn on at a setting of 1 2 on the knob Verify operation of Polarity Reversal switch applies to units made after 1 1 2005 Switch in REV position will reverse polarity at...

Page 7: ...t Nominal readings appear below for reference Normal Polarity Normal Ground 5 uA Reverse Polarity Normal Ground 5 uA Normal Polarity Open Ground 10 uA Reverse Polarity Open Ground 10 uA Measure patien...

Page 8: ...pproximately 50 DCma at this this setting which will require a resistor rated at 2 5 Watts or higher Accept 40 58 DCma Verify meter reading on 0 10 scale With a 1K load resistance set the METER SCALE...

Page 9: ...Resistor 1 00 38 0004 10K 2W 53C3 Pot 1 00 41 0013 100uf 100V Electrolytic 1 00 41 0015 100uf 250V Electrolytic 2 00 41 0033 33uf 25V Tantalum 1 00 45 0015 005uf 50V Ceramic Disc 1 00 45 0006 001uf A...