2
.
95DISP-1 ONLY
1. Plug the 95DISP-1 Display cable into the RJ
receptacle in the 95DSS3.
2. Position the 95DISP-1 Display in the
95DSS3, making sure that the notch in the
rear of the 95DISP-1 aligns with the key in the
95DSS3 housing.
3. Install the second o-ring into the machined
groove on the outside of the 95DISP-1 (not
shown)
4. Install the threaded Locking Ring onto the
95DSS3 scanner.
The locking Ring must be very firmly torqued
for the seal to be effective. A strap wrench or
similar tool with a 12 inch minimum handle is
recommended rather than installation only by
hand
4. The displays for the
InSight II are slotted in four
(4) places to mate up with
the key on the scanner and
hold the display in position
when the locking ring is
removed..
CAUTION - Damage to electronic components through electrostatic discharge (ESD)
Note: Before touching the internals of the InSight II scanner installers can discharge any static built
up on their body by touching the outside of the scanner housing if the ground is already attached. If
the scanner isn’t already grounded, the installer could touch a nearby object that is earthed.