13
ENG
Electrical connection
7.2 Local installation
The sintering oven requires its own electrical circuit.
The building's electrical circuit must have a type-K or type-Z circuit breaker with a rated current of at least
16 A (other types of circuit breaker depending on the country of use).
An additional residual-current circuit breaker (designed for 30 mA tripping current) must be installed.
To ensure electrically safe operation, the sintering oven requires a protective earth conductor connected
to the power socket.
When selecting the installation location, ensure that the accompanying mains cable is 2.0 m long. Ex-
tending the cable is not permitted. The supply voltage must be within the rated voltage range of 200 - 240
V (see section „Technical Data“).
WARNING
Release of pollutants!
Pollutants can be inhaled when the insulating materials are handled.
➤
Wear personal protective equipment (respiratory protection) during disposal.
DANGER
Electricity!
Risk of death from electric shock.
➤
Do not touch live cables and components with wet hands.
➤
Observe the accident-prevention regulations when working with electric cur-
rent.
➤
Only connect the unit to a voltage supply that matches the specifications on
the rating plate.