background image

Revised 08/03

6

APPLICATION NOTES FOR TGS2611

2-6 

Preheating of final assembly

To stabilize the detector assembly before gas
testing, the minimum period for preheating final
assemblies should be 48 hours at room
temperature (20~25˚C). Be certain to maintain
clean atmospheric conditions for preheating.

2-7 

Gas test

Test all finished products in the target gas under
normal operating conditions. Keep the
atmospheric conditions in the chamber stable,
utilizing a user-defined standard test condition
which is based on applicable performance
standards and on anticipated usage for
detectors. Remove any traces of smoke,
adhesives, gases, or solvents from the chamber.

NOTE: Without testing after final assembly,
detectors have no guarantee of accuracy or
reliability.

2-8 

Storage of finished products

Detectors should be stored in a clean air environ-
ment at room temperature.  Avoid storage in
dirty or contaminated environments. Cautions
listed in Sec. 6-1.3 of “General Information for
TGS Sensors” should also be observed.

3. Anticipated Performance at 10%LEL of

Methane

When using the classified TGS2611 with Figaro’s
recommended R

L

 for 10%LEL (Table 1) and

temperature compensated circuit design (Figure
2), typical alarm tolerances for 10%LEL of
methane such as those shown in Figure 9 are
expected. Each R

L

 classification contains a range

of tolerance as exemplified by the alarming
range labelled as ‘standard conditions’ (i.e. these
conditions are well-controlled). When factoring
in the additional effects of environmental
extremes and the allowable variation in circuit
conditions, the resulting alarming range would
be typified by the range labelled as ‘operating
conditions’. However, in actual usage, alarm
thresholds may vary since the threshold is also
affected by factors such as the tolerances of the
thermistor and/or other components, load
resistor value, test conditions, and heat
generation inside the detector enclosure.  As a

Figure 9 - Expected performance of methane detectors

with classified TGS2611 & recommended R

L

 for 10% LEL

(

refer to Table 3 for test conditions

)

d

n

a

e

r

u

t

a

r

e

p

m

e

T

y

t

i

d

i

m

u

h

d

r

a

d

n

a

t

S

s

n

o

i

t

i

d

n

o

c

0

2

±

5

6

,

C

˚

2

±

H

R

%

5

g

n

i

t

a

r

e

p

O

s

n

o

i

t

i

d

n

o

c

H

R

%

5

9

~

0

3

,

C

˚

0

4

~

0

1

-

t

i

u

c

r

i

C

n

o

i

t

i

d

n

o

c

d

r

a

d

n

a

t

S

s

n

o

i

t

i

d

n

o

c

0

.

5

=

c

V

±

C

D

V

1

0

.

0

0

.

5

=

H

V

±

C

D

V

5

0

.

0

g

n

i

t

a

r

e

p

O

s

n

o

i

t

i

d

n

o

c

0

.

5

=

c

V

±

C

D

V

2

.

0

0

.

5

=

H

V

±

C

D

V

2

.

0

r

o

i

r

p

g

n

i

n

o

i

t

i

d

n

o

C

t

s

e

t

o

t

s

r

u

o

h

8

4

Table 3 - Test conditions for measuring

performance of methane detectors as shown in Figure 9

3

4

5

6

7

8

9 10

20

15

30

under std

test conditions

under operating conditions

at recommended circuit condition

Alarming point (%LEL)

25

Summary of Contents for TGS2611

Page 1: ...bly 5 FinalAssembly 6 Preheating of Final Assembly 6 Gas Test 6 Storage of Finished Products 6 Anticipated Performance at 10 LEL of Methane 6 Pre calibrated Sensor Module 7 Appendix 8 See also Technic...

Page 2: ...er to the brochure General Information for TGS Sensors for more details Since the ID number corresponding to sensor resistance in methane gas is indicated on the sensor cap the load resistor value can...

Page 3: ...can be considered to have occurred when VRE 0 2V typ drops to near 0V Please note that a circuit voltage Vc of 5 2V should be applied to a circuit which incorporates a heater malfunction detection cir...

Page 4: ...w the lower end of the threshold range thus eliminating frequent intermittent alarming 1 7 Alarm prevention during warm up As described in Sec 2 6 of Technical Information for TGS2611 when energizing...

Page 5: ...inants excessive dust solvent vapors etc Room temperature should be 20 25 C 2 2 RL selection Choose the proper resistor for RL by referring to the ID number of the sensor and Table 1 2 3 PCB assembly...

Page 6: ...ch as those shown in Figure 9 are expected Each RL classification contains a range of tolerance as exemplified by the alarming range labelled as standard conditions i e these conditions are well contr...

Page 7: ...ld after final assembly detectors have no accuracy or reliability guarantee FIGARO GROUP HEAD OFFICE Figaro Engineering Inc 1 5 11 Senba nishi Mino Osaka 562 8505 JAPAN Tel 81 72 728 2561 Fax 81 72 72...

Page 8: ...i d G 6 6 5 5 S a b i h s o T R 1 r o t s i s e R 7 5 3 W 2 1 1 R 2 1 r o t s i s e R k 0 1 W 8 1 5 D 4 n o i t a c i f i t c e R e d o i d G 6 6 5 5 S a b i h s o T R 2 r o t s i s e R k 5 1 1 W 8 1...

Reviews: