FIBRO 68-76-00-0001 Operating Instructions Manual Download Page 28

|   28

For corona treated polymers water can be used as the single liquid to 
determine SFE in accordance with ASTM D5946-0. The SFE result is then 
presented directly in the Static window when the ASTM parameter in the 
Options Menu is set to “ON”(see Section 9.11).  

For determination of SFE using other liquids please see Section 11 of this 
User’s Manual.

Please note SFE calculations can only be done from static contact angle test 
results!
 

8.9 

Options screen

8.9.1  Enhance Mode

When a water droplet is applied on a surface of low reflectance (e.g. dark, 
rough craft liner), the droplet base will appear against a dark background. 
If the light level is increased, or the parameter CameraGain /Threshold is 
adjusted to distinguish the droplet base from its background, the droplet 
top can become too bright against the illuminated background. This is 
where the image enhancing routine will be very helpful. The enhancing 
routine is available in Static Mode, Dynamic Mode and Advancing/Receding 
Mode. Please note the image enhancing function remains in operation also 
when a new test mode is selected (e.g. changing from Static to Dynamic 
Mode). 

a) 

Double click on the labels Offset Gain, Offset Exposure and  

 

Enhance to start with the default settings.

b) 

Place the instrument on the test surface and apply a liquid  

 

droplet on the surface.

 
c) 

Double-click in the video image to switch from greyscale to a  

 

binary image if necessary.

Adjust the Camera Gain control until the top of the droplet is completed. 
Then use the Enhance control to create a solid black contour all the way 
down to the base of the droplet. 

8.9.2  Base Line determination 

A water droplet applied on a reflecting background will appear as a dark 
’saucer’ when the contact angle is below 90 degrees. A high contact angle 
above 90 degrees will appear as a ’black snowman’ with a waist against 
the white background. Similarly a contact angle of 90 degrees will result 
in a black circular shape. All these droplet silhouettes will be analyzed 
automatically and the program will assign a base line at the interface 
between the droplet and the test surface. The determined base line is 
displayed as a dotted line in the captured image. The base line can be 
adjusted manually as described.
By definition, the base line is the lowest section of the droplet resting on top 
of the test surface.

8.9.3  Camera Gain and Threshold  

The default value for Camera Gain and Threshold is established during 
calibration. This light level determines if a pixel is black or white in the binary 
image. These values can be modified by the operator to manage unusually 
bright or dark test surfaces as part of a test profile stored under a button on 
the Options screen

8.9.4  Field of View limitations

The measurement of very low contact angles is limited by the horizontal field 
of view. 
A standard PGX+ instrument has a horizontal view of approximately 6.5 
mm. Droplets of 0.5 µL will reach a width of 3.7 mm at a contact angle of 
six degrees. At an angle of 2.5 degrees this droplet will have a width of 
five mm. In comparison a 0.25 µL droplet will reach a width of 4.6 mm at a 
contact angle of 1.5 degrees. As we cannot expect the droplet to be perfectly 
centered in the image, a useful width is some 4.5 mm at the droplet base. 
This means 0.5µL droplets can be measured down to 3 degrees.  Similarly, a 
0.25 µL droplet can be measured down to a contact angle of 1.5 degrees. For 
even lower contact angles, the droplet volume has to be decreased further.

Summary of Contents for 68-76-00-0001

Page 1: ......

Page 2: ......

Page 3: ...ng the eventual transport protections Before shipping an authorization to return should be secured from Fibro System by telephone FAX or written letter The best way to ship will also be advised at tha...

Page 4: ...4...

Page 5: ...ial specimens 13 6 3 Pump dispenser 14 7 Software functions 15 7 1 Graphmaster camera 15 7 2 Graphmaster test 16 7 3 Graphmaster database 20 8 Test methods 21 8 1 Static Mode 21 8 2 Dynamic mode 22 8...

Page 6: ...6...

Page 7: ......

Page 8: ...ing the instrument is connected to the special PC software The software gives you an easy overview of the possible tests During the test the data transferred to the computer The pictures taken and the...

Page 9: ...sta Windows 7 Windows 8 Windows 8 1 or Windows 10 3 3 Scope of supply Your PGX instrument is delivered in a carrying case see picture including PGX software one millilitre syringe calibration plate gl...

Page 10: ......

Page 11: ...and I follow the displayed instructions When the installation is terminated the new device 68 76 Pocked Goniometer PGX is ready to use Double click on the device to start the Test software Note If on...

Page 12: ...the PGX instrument on top of the calibration plate with its reference mark positioned above the dot as indicated in the picture Push gently on top of the instrument to make sure the instrument rests f...

Page 13: ...e g tissues textile and film or warping materials e g newsprint cigarette tipping paper should be mounted on top of a flat backing material using a soft sticky adhesive material A Place the PGX instru...

Page 14: ...the pump sequence to inhibit application of the droplet The liquid system is filled in the following way a Place the instrument on top of a flat surface b Locate the tiny air hole on top of the white...

Page 15: ...ize is adjustable to not interfere with other windows Double click on the image to switch the live image between greyscale and binary black white mode Click with the right button on the image to open...

Page 16: ...it It allows to close the Camera window even if the software doesn t reply 7 2 Graphmaster test This manual covers only the functions specific for the 68 76 Pocket Goniometer see the GraphMaster manua...

Page 17: ...ange is always available even when the test substrate appears too dark or too bright due to variations in light reflectance Double click on the label Gain Offset to restore the default value 24 Exposu...

Page 18: ...Down keys or pick up the base line handle 1 with the mouse and slide it to the correct position Adjust the position of droplet contour When the automatic contour is set incorrectly keep press Shift a...

Page 19: ...t Defines the minimum height of a droplet reduce for smooth surfaces Default value 10 Sobel Filter Enable the Sobel filter to improve measurements on rough surface but it can increase the stability of...

Page 20: ...7 2 5 Calibration 7 3 Graphmaster database 7 3 1 Surface energy The Surface Energy functions is available only for the Static and Dynamic mode The function is enabled only when 2 or 3 records are sele...

Page 21: ......

Page 22: ...ond impossible to measure Instead the droplet is gently transferred to the surface from a very short free fall distance 8 2 1 Parameters Auto Method Off Manual test Volume Automatic test with droplet...

Page 23: ...ing tip must be equipped with a fine capillary tube inserted into the PTFE tubing P N 860309 In addition the distance between the dispensing tip and the test surface must be short enough to make sure...

Page 24: ......

Page 25: ......

Page 26: ...e2 are the two times enter as parameters and the Angle1 and Angle2 are the angle measures at the prefix times The end test will be the values of the second time 8 6 1 Parameters Auto Method Off Manual...

Page 27: ...with the manual results superimposed Use the slider below the image to shift to another image 8 8 Surface free energy SFE As this surface property cannot be measured directly it is calculated from co...

Page 28: ...let applied on a reflecting background will appear as a dark saucer when the contact angle is below 90 degrees A high contact angle above 90 degrees will appear as a black snowman with a waist against...

Page 29: ......

Page 30: ...he pump lid screw C is firmly tightened c If still no liquid is pumped release stop screw B a turn and then tighten the pump lid screw C firmly again d Repeat c above until liquid is pumped inside tub...

Page 31: ...d operating systems as these are completely outside our control Install a licensed Microsoft version on your computer re install the PG software and the problem should be gone Jumping image With a lim...

Page 32: ...in contact angle testing for historical reasons In general the contact angle is not affected by the droplet size as long as the droplet is not affected by the gravity For water having a high surface...

Page 33: ...Adjustment of Pump Volume The micro pump delivers droplets in steps of 0 5 l and for a target volume of 4 0 l the pump will make eight pump movements The Static Mode can be used to measure the average...

Page 34: ......

Page 35: ...35...

Page 36: ...36...

Reviews: