FIBRO 68-76-00-0001 Operating Instructions Manual Download Page 27

27   |

8.7 

Manual testing in static and dynamic mode

In some special situations, it might be necessary to make a manual 
determination of the droplet. 
One example is when the captured droplet image is too bright and therefore 
does not have a solid contour resulting in a “zero result”. 
Another example is when the droplet base is hidden against a too dark 
surface and the base line is then incorrectly set. In these situations the 
captured image sequence can still be used for a semi-automatic or manual 
approach described later in this section.

Double-click on a data point in the diagram or a line in the Results Window 
to bring up the corresponding image as shown in Figure 9.9. 

8.7.1  Adjust the image contrast

Double-click the image to select binary mode and slide the handle Gain 
offset up or down to make the binary image brighter or darker. Try to find 
a position where the contour is continuous and the base of the droplet 
separated from the background. This will activate the button Save, with its 
light indicator. Right-click on this button to assign a new mode (Fig 9.8). If the 
button already has the appropriate selection (e.g. “All images”), simply click 
the button to change the contrast settings for selected images. These images 
will then be recalculated using the new gain setting. Use the slider below the 
image to step between images and check the results.

If a suitable contrast setting cannot be found using the Gain offset control, it 
will be necessary to use the Enhance function.. (see section 8.9.1
 

8.7.2  Adjust the position of an incorrect baseline

When the automatic base line is set incorrectly, pick up the base line handle 
(BL) in Figure 9.9 with the mouse and slide it to the correct position. This 
will activate the button Save data with its light indicator. Right-click on the 
button to assign a new mode. If the button already has the appropriate 
mode (e.g. “All images”), simply click the button to move the base line for 
selected images. These images will then be recalculated with the new base 
line setting. 

8.7.3  Exclude fragments above droplet

When the automatic analysis picks up fragments of the dispensing tip, 
slide the top line handle (TL) below the fragments that could disturb the 
measurement. This will activate the button Save data and the changed 
setting will update all images in the measurement. The new top line setting 
will also update the region setting of the corresponding test button on the 
Image Control Panel.

8.7.4  Make a manual measurement

Select the greyscale view and click the button Manual to open the zoom 
window (Figure 9.10).
Click with the cursor at the contact point between the droplet base and the 
test surface. 
Click again at the contact point on the opposite side and once at the top of 
the droplet.
The manual results will now appear in the top left-hand corner of the 
image together with the original data. To save the manual results instead of 
the original results, click on “Save data”, which will also store the manually 
selected points superimposed in the image. To select three new points along 
the contour, click on “Reset”. To close the Zoom window, click on “Close”. 

Save Image stores a copy of the image with the manual results 
superimposed. 

Use the slider below the image to shift to another image.

 
 

8.8 

Surface ‘free’ energy (SFE)

As this surface property cannot be measured directly; it is calculated from 
contact angle measurements using different, properly characterized liquids. 
Depending on the substrate, one or more liquids are used. Please note SFE 
calculations are subject to the following constraints:

• 

No penetration of test liquid into surface (check that volume 
remains constant over time)

• 

No chemical reaction between test liquid and surface (check con-
tact angle is constant over time)

• 

Surface must by sufficiently smooth

Here the operator starts by selecting a Calculation Model from the following 
options:

• 

ASTM D5946

• 

Polar/Dispersive

• 

Acid/Base

The first option, ASTM D5946 is the “Standard Test Method for Corona-Treated 
Polymer Films using Water Contact Angle Measurements” (copyright ASTM 
International) where the water contact angle is converted into a dyne value. 
For this option only the topmost liquid field (Water) is activated and used for 
the SFE calculations.
 
The second option ‘Polar/Dispersive’ uses contact angle readings from two 
selected test liquids to determine the SFE of a substrate. Here the two first 
liquid fields (Water and DIM) are activated.
For these calculations, the Harmonic approach (Wu) is used as the first option 
as this is considered ‘more accurate’. This approach can, however, end up in 
negative roots, so there is not always a solution to these equations. In this 
situation the program changes to the Geometric approach (Owens-Wendt), 
which always gives a solution. The SFE report indicates which method was 
eventually used (Harmonic or Geometric) for the calculation. It is possible to 
force the Geometric method if the parameter SFEMETHOD is changed from 
H to G (see section 11). 

The third option ‘Acid/Base’ (Lewis) uses three liquids to calculate the surface 
energy from an equation system. Here all three liquid fields are activated 
(Water, DIM and Formamide). 
Please note the calculated results depend on the liquid combination used. 

To select an input data file, use the associated Browse button.
For Water, the browse button will present files with a wgs extension only as 
these should contain results from water contact angles. In a similar way the 
other two browse buttons (if active) can be used to select other result files 
measured with DIM or Formamide. As soon as the result file for Water has 
been selected, the first part of the file name is used for the field ‘Sample ID’ 
(e.g. 45901). 

The program will automatically search for associated result fields. The 
condition is, however, that all files must use the same ID as a file name (e.g. 
45901.wgs and 45901.dgs). The operator can still overwrite information 
displayed in the contact angle fields, in which case the Sample ID window 
will be erased. As soon as the input fields (one, two or three) have been 
completed, the program will automatically calculate the SFE results. When 
manual data is typed into the input fields, it is necessary to click on ‘Calculate’ 
to determine the SFE values. The SFE results are displayed in the Results 
Window at the bottom of the screen.

The Save button will store a copy of the report in the default directory (e.g. 
\fibro\pg) with the extension sfe (e.g. 45901.sfe). When the file has been 
stored the ‘Save’ button becomes inactive until a new result is calculated.

Summary of Contents for 68-76-00-0001

Page 1: ......

Page 2: ......

Page 3: ...ng the eventual transport protections Before shipping an authorization to return should be secured from Fibro System by telephone FAX or written letter The best way to ship will also be advised at tha...

Page 4: ...4...

Page 5: ...ial specimens 13 6 3 Pump dispenser 14 7 Software functions 15 7 1 Graphmaster camera 15 7 2 Graphmaster test 16 7 3 Graphmaster database 20 8 Test methods 21 8 1 Static Mode 21 8 2 Dynamic mode 22 8...

Page 6: ...6...

Page 7: ......

Page 8: ...ing the instrument is connected to the special PC software The software gives you an easy overview of the possible tests During the test the data transferred to the computer The pictures taken and the...

Page 9: ...sta Windows 7 Windows 8 Windows 8 1 or Windows 10 3 3 Scope of supply Your PGX instrument is delivered in a carrying case see picture including PGX software one millilitre syringe calibration plate gl...

Page 10: ......

Page 11: ...and I follow the displayed instructions When the installation is terminated the new device 68 76 Pocked Goniometer PGX is ready to use Double click on the device to start the Test software Note If on...

Page 12: ...the PGX instrument on top of the calibration plate with its reference mark positioned above the dot as indicated in the picture Push gently on top of the instrument to make sure the instrument rests f...

Page 13: ...e g tissues textile and film or warping materials e g newsprint cigarette tipping paper should be mounted on top of a flat backing material using a soft sticky adhesive material A Place the PGX instru...

Page 14: ...the pump sequence to inhibit application of the droplet The liquid system is filled in the following way a Place the instrument on top of a flat surface b Locate the tiny air hole on top of the white...

Page 15: ...ize is adjustable to not interfere with other windows Double click on the image to switch the live image between greyscale and binary black white mode Click with the right button on the image to open...

Page 16: ...it It allows to close the Camera window even if the software doesn t reply 7 2 Graphmaster test This manual covers only the functions specific for the 68 76 Pocket Goniometer see the GraphMaster manua...

Page 17: ...ange is always available even when the test substrate appears too dark or too bright due to variations in light reflectance Double click on the label Gain Offset to restore the default value 24 Exposu...

Page 18: ...Down keys or pick up the base line handle 1 with the mouse and slide it to the correct position Adjust the position of droplet contour When the automatic contour is set incorrectly keep press Shift a...

Page 19: ...t Defines the minimum height of a droplet reduce for smooth surfaces Default value 10 Sobel Filter Enable the Sobel filter to improve measurements on rough surface but it can increase the stability of...

Page 20: ...7 2 5 Calibration 7 3 Graphmaster database 7 3 1 Surface energy The Surface Energy functions is available only for the Static and Dynamic mode The function is enabled only when 2 or 3 records are sele...

Page 21: ......

Page 22: ...ond impossible to measure Instead the droplet is gently transferred to the surface from a very short free fall distance 8 2 1 Parameters Auto Method Off Manual test Volume Automatic test with droplet...

Page 23: ...ing tip must be equipped with a fine capillary tube inserted into the PTFE tubing P N 860309 In addition the distance between the dispensing tip and the test surface must be short enough to make sure...

Page 24: ......

Page 25: ......

Page 26: ...e2 are the two times enter as parameters and the Angle1 and Angle2 are the angle measures at the prefix times The end test will be the values of the second time 8 6 1 Parameters Auto Method Off Manual...

Page 27: ...with the manual results superimposed Use the slider below the image to shift to another image 8 8 Surface free energy SFE As this surface property cannot be measured directly it is calculated from co...

Page 28: ...let applied on a reflecting background will appear as a dark saucer when the contact angle is below 90 degrees A high contact angle above 90 degrees will appear as a black snowman with a waist against...

Page 29: ......

Page 30: ...he pump lid screw C is firmly tightened c If still no liquid is pumped release stop screw B a turn and then tighten the pump lid screw C firmly again d Repeat c above until liquid is pumped inside tub...

Page 31: ...d operating systems as these are completely outside our control Install a licensed Microsoft version on your computer re install the PG software and the problem should be gone Jumping image With a lim...

Page 32: ...in contact angle testing for historical reasons In general the contact angle is not affected by the droplet size as long as the droplet is not affected by the gravity For water having a high surface...

Page 33: ...Adjustment of Pump Volume The micro pump delivers droplets in steps of 0 5 l and for a target volume of 4 0 l the pump will make eight pump movements The Static Mode can be used to measure the average...

Page 34: ......

Page 35: ...35...

Page 36: ...36...

Reviews: