Update the values
Once the contact angle is measured in the correct position, press one of the
buttons (3) to re-calculate the measurement with the new settings.
The dimension and position of the contour is used only for the selected
image, the baseline position can be used to recalculate all the images or only
the forward images.
Fix baseline position
If the option “Fix” is selected the new baseline position will be used for all the
future tests.
Use the slider below the image (2) to shift to another image.
7.2.4 Hardware settings
Hardware Setting allows to customize the software
Minimum height
Defines the minimum height of a droplet (reduce for smooth surfaces)
Default value = 10
Sobel Filter
Enable the Sobel filter to improve measurements on rough surface but it can
increase the stability of the readings.
Default = false
Auto Detect
Defines the distance to the trigger line in Dynamic Mode
Default = 25
Seek Area
Seek Range for Advancing/ Receding function
Default = 40
Low Gain
Use a low gain in the SuperLow method
Default = false
Correction
Correction method in surface tension procedure.
Default = true
Curtain
The curtain allows to avoid to analyze the top part of the image.
The value is expressed in pixel.
Default = 60
Always Show Volume
It shows always the volume of the droplet.
Default = false
Hanging Time
Delay from the end of the pumping operation and the release of the droplet.
Default = 1