Display
F1X5_RS
Data: 13/01/16
file: F1X5_RS_ing.doc
Pag 3/12
Description
The F1X5_ is a mono-bidirectional microprocessor multiple keyboard counter, that processes signals
supplied by incremental transducers (encoders, optical lines, etc…) on a 5 digit display (reading scale –99999
+99999). The programming of this microprocessor is carried out by means of four keys on the frontal panel,
after introducing a password. The microprocessor allows the operator to set the input pulse correction
factors, and to multiply these factors by a 6 digit coefficient. The device has a serial output connection RS485
(or RS232) with MODBUS RTU protocol to dialogue with remote units. This instrument can keep its data in
memory also when the machine is switched off thanks to a not volatile Eeprom memory. The Reset-Enter
digit is employed to input a Preset quote too. By means of the input located on the terminal-board you can
switch on the Reset or Preset functions. This instrument is set inside a 48 x 96 panel case according to DIN
43700.
Front view
RESET
ENTER
PGM
set1
set2
Rear view
1 2 3 4 5 6 7 8
1 2 3 4 5 6
Terminal Block
Power Supply/Transducer
Terminal Block:
Serial Output
M2
M1
7 8
9 10
Installation
Before installing the instrument, read the following warnings:
a) Connect the instrument strictly following the instructions of the manual.
b) Carry out the connections using the correct wires within the limits of the tension and power supply as indicated
in
the technical data.
c) The instrument does not have an ON/OFF switch, hence it comes on when connected to the power supply. For
safety reasons, the equipment connected permanently to the power supply requires a bi-phasal selector switch
which should be within easy reach of the operator.
d) If the instrument is connected to any apparatus not isolated electrically, carry out an earth connection to avoid it
being connected directly through the structure of the machine.
d) It is the responsibility of the user to check, before using, the correct settings of the parameters of the instrument
to avoid damage to persons or things.
e) The instrument cannot function in a dangerous environment (inflammable or explosive). It can be connected to
elements that operate in the same atmosphere only through appropriate interfaces, according to the current safety
regulations.
g) Avoid dust, humidity, corrosive gases, heat sources.
h) Do not block air vents, the work temperature should be maintained between 0 ÷ 50°C.