L011-61
Page 22 of 25
2019-08-08
Rev. F0
32. Confirm the lead location.
33. Loosen and raise the lead insertion tube out of the skull.
34. Secure the lead next to the skull with smooth tweezers or the selected securing device like Medtronic Navigus cap.
35. Remove the stylet from the lead.
36. Loosen the lead holder thumbknob only slightly, ensuring that it still retains the LeadLoc and that the LeadLoc cannot move
vertically.
37. Pinch the LeadLoc and pull the lead down through the drive.
38. Remove the mounted equipment and the platform.
WARNING: Trying to pull the LeadLoc over the lead may dislodge the lead from its implantation location.