Pre-Use Check
1. Confirm that there are no contaminants or debris on any of the parts.
2. Confirm all thumbknobs are present. Note that a sterile marking pen may be used to mark the screw heads prior to use,
making it easier to differentiate between them during the surgical procedure.
3. Test mount the lead holder on the inserter, confirm there are no stripped threads
or looseness.
4. Pull the stylet up 2cm out of the insertion tube. Ensure that the stylet tip retracts
into the insertion tube.
5. Place the center hub into the integrated indexing ring. Secure the hub with the
two indexing ring screws. Ensure correct fit and seat of the hub within the ring.
Repeat for the second side for a bilateral platform.
WARNING: Delicate insertion tubes should always be carefully inspected
before use because damage to them can cause targeting errors and impact
patient safety. Handle insertion tubes carefully to prevent bending.
8mm
25mm
25mm
25mm
8mm
15mm
25mm
15mm