FHC 66-EL-LP Directions For Use Manual Download Page 6

L011-57 

   Rev. 

D1 

   2018-11-09 

   Page 

of 

11

Macrostimulation:   

From macro-contact to common or between 

macro contacts

•  Range: 0 to 10.0 mA, Display Resolution: 10 μA, 

Accuracy: ±5%, see performance graphs

Microstimulation:  

Stimulate from microelectrode to common 

•  Range: 0 to 100μA, Display Resolution: 1 μA,  

Accuracy: ±5%, see performance graphs

Stimulus Parameters

•  Monophasic or biphasic, cathode or 

anode leading or user defined

•  Pulse Train Frequency (Hz): 1 - 300 

•  Pulse Duration (µs): 20.8 - 3000 

•  Train Duration: Manual or timed (s): 0.5, 1, 

2, 3, 5, 10, 20, 30, 60

Stimulation

Mode: Constant Current, Compliance Voltage: ±50V

Impedance Check 

Measurement Waveform: sine wave, Frequency: 100Hz to 5kHz (1kHz default), Amplitude 1.0μA, Duration 1.0 to 10.0 s adjustable 
Measurement Ranges: auto-ranging 10M, 1M, 100k, 10k 
Accuracy: ±10% of full range, Resolution: 2 digits

Power Requirements

 

Input: 100 -240 VAC, 50 /60Hz. 
Fuse: 2A 250V Type T H

Analog Inputs

Quantity – two, earth referenced, patient-isolated, AC/DC Coupled, Bandwidth 0-5kHz 
Ranges (factory configurable): 0 to 16V, -8.0V to +8.0V, 0 to 8.0V, -4.0V to +4.0V, 0 to 1.6V, -800mV to +800mV, 0 to 800mV, -400mV to 
+400mV, 0 to 160mV or -80mV to +80mV

Inventory:

MT-LPP:  Guideline 4000 LP+

1.  Notebook computer
2.  microTargeting Controller module (sold separately)*
3.  microTargeting Controller remote control (sold separately)*
4.  microTargeting Controller motor or encoder assembly  

(sold separately)*

5.  Inline serial–to-USB converter (sold separately)*
6.  Stimulation remote control
7.  Speakers
8.  Audio cable
9.  microTargeting electrode cable for LP+ (sold separately)**
10.  Power adaptor for notebook computer (1)
11.  Power adaptor for speakers (7)
12.  Power cord for Guideline 4000 LP+ 
13.  Power adaptor for microTargeting Controller  

(sold separately)*

14.  Power strip, max load 10A 240V
15.  Power cord for power strip (multiple portable  

socket outlets; connect to mains)

16.  Guideline 4000 LP+ test electrode
17.  Breakout box (sold separately)***
18.  USB cable
19.  USB-powered optical mouse

* Included with 66-DS-PA or 66-DS-PD kits, ** Sold as 66-EL-LP, *** Sold as MT-LPP-BOX

1

3

4

2

5

6

7

8

9

10

11
12

13

14

15

16

17

18

19

Summary of Contents for 66-EL-LP

Page 1: ...products MT LPP MT LPP APM 66 EL LP FHC Inc 1201 Main Street Bowdoin ME 04287 USA Fax 1 207 666 8292 www fh co com FHC Europe TERMOBIT PROD srl 42A Barbu Vacarescu Str 3rd Fl Bucharest 020281 Sector 2...

Page 2: ...L011 57 Rev D1 2018 11 09 Page 2 of 11...

Page 3: ...y Warnings and Cautions System Maintenance and Repair Handling Considerations Maintenance Repair System Disposal System Specifications Performance Accuracy Microelectrode Recording Stimulation Impedan...

Page 4: ...complies with MDD Directive 93 42 EEC and legal responsibilities as a manufacturer are with FHC Inc 1201 Main Street Bowdoin ME 04287 USA Temperature limitation Storage conditions are between 0 C 40 C...

Page 5: ...osal Return the Guideline 4000 LP to FHC for proper end of life disposal Please contact an FHC factory authorized representative to arrange a return System Specifications Performance Accuracy Microele...

Page 6: ...patient isolated AC DC Coupled Bandwidth 0 5kHz Ranges factory configurable 0 to 16V 8 0V to 8 0V 0 to 8 0V 4 0V to 4 0V 0 to 1 6V 800mV to 800mV 0 to 800mV 400mV to 400mV 0 to 160mV or 80mV to 80mV...

Page 7: ...speakers or Guideline 4000 LP and the patient simultaneously If using headphones use wireless headphones if contact with patient is possible WARNING The Guideline 4000 LP Multiple Portable Socket Out...

Page 8: ...nfirm that a scrolling waveform is visible in the corresponding thumbnail window for each installed channel a If any channels have failed to initialize properly wait one minute for the LP to detect an...

Page 9: ...nect the test electrode before connecting the LP to a patient 4 Load a user profile and enter patient information a Remove the test electrode if it is still attached to the LP b Click on the Patient t...

Page 10: ...n using the orange activate button on the stimulation remote control If desired stimulation may be stopped by pressing the orange activate button again during fixed duration mode only WARNING Electrod...

Page 11: ...e top center of the screen to view root mean square RMS voltage versus depth and mean firing rate MFR versus depth plots for all channels d Click on the Guidance tab at the top left of the computer sc...

Reviews: