System Maintenance and Repair:

Handling Considerations

•  The Guideline 4000 LP+ is intended to be used in a standard operating room environment.
•  Store the Guideline 4000 LP+ at room temperatures between 0°C (32°F) and 40°C (104°F).
•  Neither the Guideline 4000 LP+ nor its components are designed for sterilization. 
•  Do not drop the Guideline 4000 LP+ or its components. 
•  Only transport the Guideline 4000 LP+ inside of its shipping container, with included foam insulation.  

 
 
 
 

Maintenance

•  The Guideline 4000 LP+ must be inspected and recalibrated each year by an FHC authorized representative.  Contact FHC for 

more information on Service Agreements and Preventive Maintenance Programs.

•  The Guideline 4000 LP+  and its components may be cleaned by wiping a dry rag lightly over all surfaces. If necessary, a slightly 

damp cloth  moistened with a standard hospital disinfectant may be used. 
 

Repair

•  All FHC products are unconditionally guaranteed against defects in workmanship for one year from the date of shipment, 

provided they have been exposed to normal and proper use. 

•  Should service or repair be required, please contact FHC at 1-800-326-2905 (US & Canada) or +1-207-666-8190 for return 

instructions.

End-of-Life System Disposal

Return the Guideline 4000 LP+ to FHC for proper end-of-life disposal. Please contact an FHC factory authorized  
representative to arrange a return.

System Specifications - Performance & Accuracy:

Microelectrode Recording

Analog signal conditioning

•  Gain: 54 to 7201 user adjustable,  

Amplitude accuracy: ±5%, Polarity: Inverting 

•   Upper frequency cutoff: 20kHz, 40dB/decade 

roll-off 

•  Lower frequency cutoff: MER - 200Hz, 20dB/

decade; LFP - <1Hz, 20dB/decade

Digital signal processing

•  Sampling: 16 bit resolution over ±1.4V range

•  Sample Rate: MER: 10-48kHz ,LFP: 5-1000Hz  

•   Digital Bandpass Filter: High Pass - 0 to 1000Hz, 

Low Pass – 100 to 20kHz 

•   Adaptive line noise filter: adaptation weight  

adjustable from 0 to 10% per cycle 

•   Adaptive Stimulus Artifact Suppressor: adaptation 

weight adjustable from 0 to 10% per cycle

•   Timing accuracy: ±10ppm (0.001%), resolution: 

1/48,000 s (~20.8µs)

Waveform Analysis

•  Display Mode: scroll/sweep rate  

adjustable from 5s/window to 0.25s/ 

 

window 

•   Spikes Mode: duration 290μs/sweep to 

2.585 ms/sweep in 12 steps 

•  Spike discrimination method: Dual line 

level or Window

•  Firing Rate: displayed in Hz, one second 

refresh rate

 Password Reminder:

WARNING: Do not expose the Guideline 4000 LP+ to humidity of greater than 95%.

WARNING: Do not store the Guideline 4000 LP+ near or expose it to strong magnetic fields.

WARNING: Do not allow any liquid to contact any Guideline 4000 LP+ component.

CAUTION: Only qualified FHC representatives may calibrate the Guideline 4000 LP+.

Summary of Contents for 66-EL-LP

Page 1: ...products MT LPP MT LPP APM 66 EL LP FHC Inc 1201 Main Street Bowdoin ME 04287 USA Fax 1 207 666 8292 www fh co com FHC Europe TERMOBIT PROD srl 42A Barbu Vacarescu Str 3rd Fl Bucharest 020281 Sector 2...

Page 2: ...L011 57 Rev D1 2018 11 09 Page 2 of 11...

Page 3: ...y Warnings and Cautions System Maintenance and Repair Handling Considerations Maintenance Repair System Disposal System Specifications Performance Accuracy Microelectrode Recording Stimulation Impedan...

Page 4: ...complies with MDD Directive 93 42 EEC and legal responsibilities as a manufacturer are with FHC Inc 1201 Main Street Bowdoin ME 04287 USA Temperature limitation Storage conditions are between 0 C 40 C...

Page 5: ...osal Return the Guideline 4000 LP to FHC for proper end of life disposal Please contact an FHC factory authorized representative to arrange a return System Specifications Performance Accuracy Microele...

Page 6: ...patient isolated AC DC Coupled Bandwidth 0 5kHz Ranges factory configurable 0 to 16V 8 0V to 8 0V 0 to 8 0V 4 0V to 4 0V 0 to 1 6V 800mV to 800mV 0 to 800mV 400mV to 400mV 0 to 160mV or 80mV to 80mV...

Page 7: ...speakers or Guideline 4000 LP and the patient simultaneously If using headphones use wireless headphones if contact with patient is possible WARNING The Guideline 4000 LP Multiple Portable Socket Out...

Page 8: ...nfirm that a scrolling waveform is visible in the corresponding thumbnail window for each installed channel a If any channels have failed to initialize properly wait one minute for the LP to detect an...

Page 9: ...nect the test electrode before connecting the LP to a patient 4 Load a user profile and enter patient information a Remove the test electrode if it is still attached to the LP b Click on the Patient t...

Page 10: ...n using the orange activate button on the stimulation remote control If desired stimulation may be stopped by pressing the orange activate button again during fixed duration mode only WARNING Electrod...

Page 11: ...e top center of the screen to view root mean square RMS voltage versus depth and mean firing rate MFR versus depth plots for all channels d Click on the Guidance tab at the top left of the computer sc...

Reviews: