The Habituation protocol can be used while sitting, lying, and it does NOT require the sensor to be on or paired to the stimulator.
This protocol should be used in the following cases:
1. The very first time that the user uses the device, to get the patient used to the sensation.
2. In patients without ambulatory capacity, as a part of the previous to standing.
3. In patients whose motor threshold is above the pain threshold. This protocol can be used to lower the motor threshold to even
-
tually generate movement comfortably.
Select the configuration and the stimulation time.
Edit the intensity by increasing the value progres
-
sively.
When a motor response is achieved, it will be
possible to "Set motor threshold" and load it later
in the movement settings configuration.
Stimulation parameters can be edited (frequency,
pulse widht) by clicking on the "Expand" button.
To start with the stimulation, click on the green
button “Play” (the configuration will be saved au
-
tomatically)..
Once the session is finished, click on the red
"Stop" button.
When going "back" in the navigation bar, a mes
-
sage will appear in which you must choose be
-
tween "Save session" or "Exit without saving".
Select configuration
Edit intensity
Play
Stop
It is highly recommended to edit the intensity with caution!
This parameter is related to the comfort and safety. Then,
it is highly recommended to increase it gradually taking the
motor threshold into account.
!
1
2
3
4
Individual scanning
All the pads will be activated
sequentially, one by one.
Couple scanning
All the pads will be activated
sequentially in pairs, alternat
-
ing flexors and extensors.
Time
Can be set:
- A specific duration editing the default value (the counter will be displayed like a
countdown).
- An undefined duration clicking on the "Loop" button (the counter will start from 0).
CONFIGURATION BAR
Intensity/motor treshold
Clicking on the "Set Motor Treshold" button, the intensity value will be saved
and later in the movement configuration screen will be possible to upload.
Set motor treshold
Stimulation parameters
Start/Stop stimulation
34 INSTRUCTIONS FOR USE
|
Fesia
Walk
HABITUATION PROTOCOL
03
FESIA PRO APPLICATION