FERNO/MAN/1121/177-R04/UK
Page
16
Cleaning, Disinfecting & Maintenance
Regular cleaning helps reduce the risk of transmitting disease and enables equipment to
function at its optimum.
Ensure that all equipment is removed from the trolley and that the Power distribution
socket is disconnected from mains supply.
Trolley should be hand-wiped using a suitable cleaning solution. For particularly grimy
cases, especially if the trolley has been in storage for a long period, it may be necessary to
first clean the trolley with a water-soluble solvent-soaked cloth.
Once cleaned, dry the trolley with a soft towel, paying particular attention to swivel and
sliding joints. If a compressed air hose is available, you may find that compressed air jets
are more effective in drying the hard-to-reach places.
Ensure no liquid enters the power distribution sockets.
Cleaning webbing restraint straps
1. Immerse the restraint webbing straps in a solution of mild soap and water. Do not
immerse any metal buckles or metal fixation clips in the solution, these should be wiped
clean only.
2. Repeatedly dip the webbing in clear water to rinse. Do not immerse any metal buckles
or fixation clips
3. Hang the restraints to air-dry. Attach only clean, dry restraints back onto to the
equipment
Disinfecting webbing restraint straps
1. Immerse or spray the restraint webbing straps following the selected disinfectant
manufacturer’s instructions for application method and contact time. Do not immerse any
metal buckles or metal fixation clips into disinfectant liquid
2. Spray the metal buckles and fixation clips with disinfectant cleaner spray, following the
disinfectant manufacturer’s instructions for application method and contact time
3. Hang the restraints to air-dry if needed. Attach only dried restraints to the equipment
Cleaning Biosafe restraint straps
1. Wipe-clean the restraints, buckles and fixing clips with mild soap and water
2. Hang the restraints to air-dry. Attach only clean, dry restraints back onto to the
equipment
Disinfecting Biosafe restraint straps
1. Spray the restraint webbing metal buckles and fixation clips with disinfectant cleaner
spray, following the disinfectant manufacturer’s instructions for application method and
contact time