FEMA ELECTRÓNICA . Series M . M40-P
3
1.2 Front view
1.3 Rear view
Earth connection - Although a
terminal is provided for earth
connection, this connection is
optional. The instrument does
not need earth connection for
correct operation nor for com-
pliance with the security regu-
lations.
~
~
+
-
8 9 0
Fuse - To comply with security regulation 61010-1, add to the power
line a protection fuse acting as disconnection element, easily acces-
sible to the operator and identified as a protection device.
Power ‘H’
fuse 250 mA time lag
Power ‘L’
fuse 400 mA time lag
1.4 Power connections
44
92
Panel
cut-out
16
8
75
Passive voltage signal 0/10 Vdc, ±10 Vdc
Active voltage signal 0/10 Vdc, ±10 Vdc
Common
Signal
Active current loop 4/20 mA, ±20 mA
Passive current loop 4/20 mA, ±20 mA
Vexc.
Signal
Common
Signal
1 2 3
4 5
Jumper 5 closed
1 2 3
4 5
Jumper 5 closed
1 2 3
4 5
Jumper 4 closed
1 2 3
Common
Vexc.
Signal
4 5
Jumper 4 closed
1.5 Signal connections
Button ‘UP’
Logo
Alarms
Units
Button ‘SQ’
Button ‘LE’
‘Configuration menu’
‘Fast access‘
1.6 Mechanical dimensions (mm)
Signal
Jumpers
Option 1
Option 2
Option 3
Power
4
1 2 3
5
8 9 0
Detail of the plug-in screw terminals provided with
the instrument. The instrument is provided with all
terminals needed, both male and female.
96
48
Units